EAST装置PKR251型真空规管的修复及标定技术研究  被引量:2

TECHNICAL STUDIES ON REPARATION AND CALIBRATION OF PKR251 VACUUM GAUGE ON EAST

在线阅读下载全文

作  者:黄明 王厚银[1] 余耀伟 左桂忠 孙震 孟献才 徐伟[1] 胡建生[1,3] HUANG Min;WANG Hou-yin;YU Yao-wei;ZUO Gui-zhon;SUN Zhen;MENG Xian- cai;XU Wei;HU Jian- sheng(Institute of Plasma Physics, Chinese Academy of Sciences,Hefei 230031, China;CAS key Laboratory of Photovoltaic and energy conservation materials, Hefei 230034, China;Department of Applied Physics, School of Physics and Elecll.onics,Hunan University, Changsha 440082, China)

机构地区:[1]中国科学院等离子体物理研究所,合肥230031 [2]湖南大学物理与微电子科学学院应用物理系,合肥230031 [3]中国科学院光伏与节能材料重点实验室,长沙410082

出  处:《真空与低温》2018年第2期92-95,共4页Vacuum and Cryogenics

基  金:国家自然科学基金(11625524、11605237、11605246、11405210、11775261)的经费支持

摘  要:EAST装置复杂的真空系统需要大量PKR251型真空规管实时测量,并获得准确的真空度数据,但是长时间运行后会出现测量结果不准的现象。PKR251型真空规管的修复和标定技术是通过对规管内部电极、极板等关键部件的去污(更换)、清洗和烘烤处理,修复计量不准或错误的真空规管。搭建规管标定平台模拟EAST装置的真空环境,通过控制充入真空室内高纯氮气的充气速率,改变真空室内的气体压力,比较单个规管与标准规的读数,判定规管测量准确度,实现规管标定。标定实验发现修复后的PKR251型规管在不同压力下线性度较好,满足EAST装置真空测量要求。In order to obtain real-time and accurate vacuum measurement, a lalge number of PKR251-type vacuum gallges have been installed on EAST tokamak with colnplex vacuum systems, but inaccurate, even wrong measured results were always observed after lollg and continuous ii111 of EAST. Tiffs paper describes a tecMique for tile repair and calibra- tion of a PKR251 vacuum gallge to restore tiffs gauges with inaccurate or wrollg measurement by decontamination (re-placement ), cleaning and baking of key components, including internal electrodes, plates. To stimulate the EAST vacmml condition, a calibration platform was built which could adjust the vacmml pressme by changing the flow rate of high-pmity nitrogen injection into the vessel of tiffs system. After the accuracy of measurements is reported by compamlg the read- ings of individual gauge and the standard, the gauge calibration is finished. It was found that the measurement of repaired gauge fitted well with the standard at different pressmes and they could be used again by EAST. This repair and calibration technology can effectively extend the service life of PKR251 gauge, supplying the techiffcal support for energy conserva- tion and environmental protection in nuclear fusion experiments.

关 键 词:EAST装置 真空测量 真空规管 修复标定 

分 类 号:TL628[核科学技术—核技术及应用]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象