摘 要:通过分析GE-MA5500心电图机2例特殊故障维修,讨论维护方法。通过分析GE-MA5500打印不停止的故障原因,查找故障源,进行维修和保养。结果显示一组数据线出现断路,导致打印信息无法回馈到控制板,引起该故障。本文描述的2例故障较为少见,尤其是第1例,需耐心查找方可确定故障源,可供其他工程技术人员参考。This paper discusses the maintenance methods of two cases of special fault maintenance of GE-MA5500 electrocardiograph.By analyzing the causes of the failure of the GE-MA5500 printing report,the failure source was found and the maintenance and maintenance was carried out.The result shows that a set of data lines was broken which causes the printing information to be unable to feed back to the control board,causing the failure.The 2 cases described in this paper are relatively rare,especially in first cases.We need to be patient reference to find out the source of failure.