GE-MA5500心电图机故障分析与维修2例  被引量:1

Two cases of failure analysis and maintenance of GE-MA5500 electrocardiograph

在线阅读下载全文

作  者:郝美硕 HAO Mei-shuo(Department of Apparatus,Tiefa Coal Group General Hospital,Liaoning Province,Tieling 112700,China)

机构地区:[1]辽宁省铁法煤业集团总医院器械科,辽宁铁岭112700

出  处:《中国当代医药》2018年第12期149-151,共3页China Modern Medicine

摘  要:通过分析GE-MA5500心电图机2例特殊故障维修,讨论维护方法。通过分析GE-MA5500打印不停止的故障原因,查找故障源,进行维修和保养。结果显示一组数据线出现断路,导致打印信息无法回馈到控制板,引起该故障。本文描述的2例故障较为少见,尤其是第1例,需耐心查找方可确定故障源,可供其他工程技术人员参考。This paper discusses the maintenance methods of two cases of special fault maintenance of GE-MA5500 electrocardiograph.By analyzing the causes of the failure of the GE-MA5500 printing report,the failure source was found and the maintenance and maintenance was carried out.The result shows that a set of data lines was broken which causes the printing information to be unable to feed back to the control board,causing the failure.The 2 cases described in this paper are relatively rare,especially in first cases.We need to be patient reference to find out the source of failure.

关 键 词:心电图机 设备维修 数据线 

分 类 号:R197.39[医药卫生—卫生事业管理]

 