作 者:赵治翔 刘白林[1] 白万民[1] 杨举 周媛 ZHAO Zhixiang;LIU Bailin;BAI Wanmin;YANG Ju;ZHOU Yuan(School of Computer Science and Engineering,Xi' an Technological University,Xi' an 710021,China)
机构地区:[1]西安工业大学计算机科学与工程学院,西安710021
出 处:《西安工业大学学报》2018年第3期293-297,共5页Journal of Xi’an Technological University
基 金:陕西省科学技术研究发展计划项目(2013K13-04-07)
摘 要:为提高医用内窥镜畸变率的检测效率,文中提出了一种基于椭圆拟合的内窥镜畸变率检测方法.对采集到的测标板上的图像进行自动阈值预处理并进行模板匹配,运用投影和反投影技术,去除多余边缘,通过直接最小二乘法进行拟合椭圆,进而求取其畸变率.文中以QM423T型号医用内窥镜为对象进行畸变率检测,选取畸变板上5个椭圆,经过20次测量,不同位置的椭圆和平均畸变率测量结果为:左上(0.897 4)、右上(0.876 5)、中心(0.934 8)、左下(0.893 0)、右下(0.907 9),测量结果均符合该型号内窥镜畸变率的测量范围,满足QM423T型号医用内窥镜的相关标准的测试要求.文中方法检测效率稳定,可重复操作,可减少人工检测的误差,对实际应用具有参考价值.A method based on ellipse fitting for detecting optical performance parameters of the endoscope distortion rate is proposed in this paper.For the image captured on the test plate,automatic threshold preprocessing and template matching were done.The excess edges were removed by the projection and back projection technology,and then the distortion rate of the fitted ellipse was calculated by the direct least square method.In this paper,the distortion rate of the QM423T model medical endoscope is detected.5 ellipses selected on the distorted board were measured 20 times.The results of the ellipse at different locations and average distortion rate are:upper left(0.897 4),upper right(0.8765),center(0.934 8),lower left(0.893 0)and lower right(0.9079).The results are within the measurement range of the endoscope distortion rate of this model,which satisfies the standards of the QM423T medical endoscope.The proposed method is stable and repeatable and is able to reduce errors in manual detection.The study is referential to its practical application.
分 类 号:TP751[自动化与计算机技术—检测技术与自动化装置]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:18.191.31.198