后装γ源近距离治疗室内辐射场剂量水平及分布  被引量:2

Dose level and distribution of radiation field in a γ-ray source after loading brachytherapy room

在线阅读下载全文

作  者:李凤霜[1] 张朝 李夏[2] 马永忠[3] LI Feng-shuang;ZHANG Zhao;LI Xia;MA Yong-zhong(Department of Gynaecalogical Oncology and Medical Engineering Center,Beijing Obetetrics and Gynecology Hospital,Capital Medical University,Beijing,100006,China;Institute for Radiation Safety,China Institute of Atomic Energy,Beijing,102413,China;Institute for Radiation Hygiene Protection,Beijing Center for Disease Control and Prevention/Beijing Center for Prevention Medicine Research,Beijing,100013,China)

机构地区:[1]首都医科大学附属北京妇产医院妇科肿瘤科和医学工程中心,北京100006 [2]中国原子能科学研究院辐射安全研究所,北京102413 [3]北京市疾病预防控制中心/北京市预防医学研究中心放射卫生防护所,北京100013

出  处:《职业与健康》2018年第15期2113-2117,共5页Occupation and Health

基  金:北京市预防医学公益应用课题(2016-BJYJ-18)

摘  要:目的掌握后装γ源近距离治疗室(简称后装治疗室)内辐射场剂量水平及其分布,为机房的防护设计和改造以及现场人员的放射防护指导提供科学决策。方法以北京市某医院1台RL-HZJ 18型192Ir后装γ源近距离治疗机(后装机)及其治疗室为研究对象,在实际放射源活度为254.7 GBq的情况下,使用AT1121型辐射巡测仪分别测量后装机贮源状态下治疗室内辐射场剂量水平和出束状态下迷路的剂量分布;使用热释光剂量计(TLD)测量患者治疗期间治疗室内关注点的辐射剂量水平。结果后装机贮源状态下距离设备表面5和100 cm处的周围剂量当量率最高值分别为24.0和0.56μSv/h;照射状态下迷路拐口周围剂量当量率为5.4×10~3μSv/h,相应的辐射比率为3.8×10^(-2),迷路内接近入口防护门处为9.4μSv/h,相应的辐射比率为6.8×10^(-5);患者治疗期间治疗室内墙面关注点的辐射比率为(0.235~201.4)×10^(-6)。结论后装治疗室迷路具有明显的防护效果,迷路的设置和治疗室内辐射剂量水平的实时监测是必要的。[Objective] To know the dose levels and distribution of radiation field in a γ-ray source after loading brachytherapy room (brachytherapy room), provide scientific decision-making for the radiation protection design and reconstruction of brachytherapy room and the guideline on radiological protection of radiotherapy onsite staff. [Methods] A 192Ir brachytherapy equipment typed with RL-HZJ 18 (brachytherapy equipment) and its room of a hospital were selected as the research objects. Under the condition of the actual source activity 254.7 GBq loaded in the equipment, the dose levels of radiation field in the room were measured with a radiation survey instrument typed AT1121 while the source being in stored state, and the dose distribution along maze were measured while the source being in exposure state. Radiation dose levels of the concerned points in the room were also measured accumulatively using thermoluminescence dosimeter (TLD) during the period of normal clinic treatment of patients. [Results] Under the condition of non-irradiation state, the maximum values of ambient dose equivalent rate around 5 and 100 cm away from the surface of the brachytherapy equipment were 24.0 and 0.56 μSv/h respectively. While the brachytherapy radioactive source was being in irradiation state, ambient dose equivalent rate measured on the maze turning place was 5.4×10^3 μSv/h which corresponding radiation ratio of 3.8×10^-5, and the ambient dose equivalent rate along the maze near entrance of the room was 9.4 μSv/h which corresponding radiation ratio of 6.8×10^-5. The radiation ratio of the concerned points on the wall surface in the room were in the range of (0.235-201.4)×10^-6. [Conclusion] It is shown that the maze of a brachytherapy room can develop obvious protective effect, and it is necessary that maze should be set and radiation levels in a brachytherapy room should be monitored in real-time.

关 键 词:后装治疗机 后装治疗室 辐射场 放射防护 

分 类 号:R148[医药卫生—公共卫生与预防医学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象