检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
作 者:李鹤[1] 王景宇[1] 赵志艺 孟阳阳 张艳娇[1] 陈亮[1] LI He;WANG Jing-yu;ZHAO Zhi-yi;MENG Yang-yang;ZHANG Yan-jiao;CHEN Liang(Department of Radiology,the First Hospital of Jilin University,Changchun 130021,Jilin,China)
机构地区:[1]吉林大学第一医院放射科,吉林长春130021
出 处:《广东医学》2018年第16期2437-2441,共5页Guangdong Medical Journal
基 金:吉林省卫生计生委科技计划(编号:20142054)
摘 要:目的探讨在保证冠状动脉CT成像质量基础上寻找合适的噪声指数(NI)以有效减少辐射剂量。方法收集临床上疑诊为冠心病患者100例,随机分为5组行CCTA扫描,应用造影剂Visipaque(270mg/mL)及100 Kev管电压扫描,分别设置NI 15.0、17.0、19.0、21.0、23.0。采用自适应迭代重建算法(ASi RV)进行重建。用统计分析软件对各组患者的辐射剂量和图像质量进行分析和比较。结果在对比剂浓度及管电压一定时,随着NI值的增加,平均容积CT剂量指数(CTDIvol)、剂量长度乘积(DLP)、有效辐射剂量(ED)逐渐降低,NI 15.0、17.0、19.0组间CTDIvol、DLP、ED差异有统计学意义(P<0.05)。NI 19.0组辐射剂量(CTDIvol、ED、DLP)较NI 15.0、NI 17.0组降低较明显(P<0.05)。噪声(SD)随着NI的增高而增高;反之,信噪比(SNR)、对比噪声比(CNR)随着NI的增高而降低,NI 17.0、19.0、21.0组间SD、SNR、CNR差异无统计学意义(P>0.05)。主动脉窦及右冠状动脉(RCA)近段、左前降支(LAD)近段、左回旋支(LCX)近段的平均CT值在5组间差异无统计学意义(P>0.05)。NI 15.0、17.0、19.0、21.0、23.0组间RCA、LAD、LCX主观评价差异有统计学意义(P<0.05)。图像质量主观评分在NI 21.0和23.0组较NI 15.0、NI 17.0和NI 19.0组降低(P<0.05)。结论 NI为19.0时,应用低浓度造影剂并结合ASiR-V算法行冠状动脉CT血管造影,其在有效降低辐射剂量的同时保证图像质量是可行的。Objective To investigate the appropriate noise index( NI) values for effective reducing the radiation dose while maintaining the quality of CT images of coronary arteries. Methods A hundred patients with clinically suspected coronary heart disease from October to December in 2017 with CCTA scanning were randomly divided into 5 groups. We adjust different value of noise index( NI) when evaluating each group every other number from 15. 0 to 23. 0,with twenty patients in each group. Contrast agent Visipaque( 270 mg/m L) was applied and tube voltage was set at100 Kev. All raw data were reconstructed with adaptive statistical iterative reconstruction-V( ASi R-V,50%). Statistical software was used to analyze and compare the radiation dose and image quality among the groups. Results When the contrast media concentration and tube voltage were fixed,as the NI value increases,the CTDIvol,DLP and ED were gradually reduced. There was a statistically significant difference among the 5 groups( P〈0. 05). The radiation dose( CTDIvol,ED,DLP) at NI 19. 0 was significantly lower than the Group NI 15. 0 and NI 17. 0. The SD increased with the increase of NI. On the contrary,the SNR and CNR decreased with the increase of NI. There was significant differences in SD,SNR and CNR among the 5 groups( P〈0. 05). There were no significant differences in the CT values of the aortic sinus and the three major coronary arteries among the 5 groups( P〈0. 05). There were significant differences in image quality subjective evaluation among the 5 groups( P〈0. 05). Image quality subjective evaluation were reduced in the Group NI 21. 0 and 23. 0. Conclusion Low concentration contrast agents and NI 19. 0 value combined with ASi R-V in coronary CT imaging reduces the dose of radiation while ensuring the image quality.
关 键 词:冠状动脉 体层摄影术 X线计算机 图像质量 噪声指数 自适应迭代重建技术
分 类 号:R541.4[医药卫生—心血管疾病] R816.2[医药卫生—内科学]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.145