杭州地区29家医疗机构全血细胞计数准确度和一致性调查  被引量:3

Survey on the accuracy and consistency of complete blood count in 29 medical institutions in Hangzhou

在线阅读下载全文

作  者:王文娟[1] 单志明[2] 武锦彪[1] 吕俊霞[1] 王星[1] 郦卫星[2] 陈瑜[1] Wang Wenjuan;Shah Zhiming;Wu Jinbiao;Lyu Junxia;Wang Xing;Li Weixing;Chen Yu(Department of Clinical Laboratory,The First Affliated Hospital,College of Medicine,Zhejiang University,Hangzhou 310003,China;Zhejiang Center for Clinical Laboratories,Hangzhou 310014,China)

机构地区:[1]浙江大学医学院附属第一医院检验科,杭州310003 [2]浙江省临床检验中心,杭州310014

出  处:《中华临床实验室管理电子杂志》2018年第3期165-167,共3页Chinese Journal of Clinical Laboratory Management(Electronic Edition)

基  金:浙江省医药卫生一般研究计划资助项目(2012KYB076);浙江省级重点科技创新团队资助项目(2010R50048-15)

摘  要:目的了解杭州地区医疗机构全血细胞计数(complete blood count,CBC)项目检测结果的准确度和一致性,为全市医疗机构间该项目检测结果的互认提供基础。方法用5份新鲜血标本对杭州地区29家代表性医疗机构进行CBC项目检测结果调查,标本同时送卫生部临床检验中心血细胞分析参考实验室定值。参照WS/T 406-2012中允许总误差要求及准确度验证方法,对回报结果进行准确度和可比性评价,及其合格率分析。结果 29家医疗机构回报了63台血液分析仪检测结果,WBC、RBC、Hb、HCT、PLT检测结果准确度验证合格率分别为100.0%、95.2%、98.4%、98.4%、100.0%;WBC、RBC、Hb、HCT、PLT、MCV、MCH、MCHC检测结果的可比性评价合格率分别为99.7%、94.9%、99.4%、100.0%、99.7%、99.7%、100.0%、100.0%。结论杭州地区医疗机构的CBC项目检测结果准确度和一致性优良,该调查活动为开展检验结果互认积累了经验,且奠定了基础。Objective To investigate the accuracy and consistency of complete blood count(CBC) in Hangzhou medical institutions in order to provide the foundation for mutual mutual recognition of the test results from different medical institutions in Hangzhou. Methods Five fresh blood samples were used to investigate CBC in 29 representative medical institutions in Hangzhou. The samples were also sent to Hematology Reference Laboratory of National Center for Clinical Laboratories. The allowable total error requirement and accuracy verification method in WS/T 406-2012 were consulted. The accuracy and comparability of the return results were evaluated, the vatio of the results that meet the requirements were qualified. Results The accuracy verification rate of the tests for WBC, RBC, Hb, HCT, PLT from the total of 63 hematology analyzers were 100.0%, 95.2%, 98.4%, 98.4%, and 100.0%. The qualified rate of the comparable evaluation of the test results for WBC, RBC, Hb, HCT, PLT, MCV, MCH and MCHC from the total of 63 hematology analyzers were 99.7%, 94.9%, 99.4%, 100.0%, 99.7%, 99.7%, 100.0%, and 100.0%. Conclusions The accuracy and consistency of CBC results are satisfied in the medical institutions in Hangzhou. This investigation has accumulated experience and foundation for the mutual recognition of test results.

关 键 词:全血细胞计数 准确度 可比性 血液 

分 类 号:R446.1[医药卫生—诊断学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象