作 者:齐文章[1] 安静[2] QI Wenzhang;AN Jing(Construction Offce;Department of Radiology,Linzi District People's Hospital,Zibo Shandong 255400,China)
机构地区:[1]临淄区人民医院建设办公室,山东淄博255400 [2]临淄区人民医院放射科,山东淄博255400
出 处:《中国医疗设备》2018年第10期110-112,122,共4页China Medical Devices
摘 要:除颤仪的正常使用是保证医院急救工作的重要部分,所以我们不仅要做好除颤仪的预防性维护工作,还要在其发生故障时,能够及时将故障排除。本文主要介绍了飞利浦M3536A除颤监护仪的三例常见故障及其排除方法,主要包括NBP测试未通过、打印机不打印、以及心电波形干扰,以期能为同行排除相似故障提供一定的参考。It is well known that the normal use of defibrillator is an important part of hospital emergency work. Therefore, we should not only do the preventive maintenance of the defbrillator, but also remove the fault quickly. Three common faults of philips M3536A defbrillation monitor and their troubleshooting methods, which mainly included NBP test failed, printer failed and ECG wave interference were introduced in this paper. We hope this paper can provide some reference for peer troubleshooting.
关 键 词:急救设备 除颤监护仪 除颤仪的日常维护 除颤仪的故障维修