飞利浦彩超硬软件的故障分析与检修  被引量:5

Failure Analysis and Maintenance of PHILPS Color Doppler Machine Hardware and Software

在线阅读下载全文

作  者:谭峰 TAN Feng(Department of Equipment,Gangbei District People's Hospital of Guigang,Guigang,537100)

机构地区:[1]贵港市港北区人民医院,贵港市537100

出  处:《中国医疗器械杂志》2018年第5期388-390,共3页Chinese Journal of Medical Instrumentation

摘  要:该文首先对飞利浦彩超机的结构及工作流程进行分析,并以飞利浦彩超M2540A的电子箱故障和飞利浦彩超HD15T的硬件和软件故障为例详细分析了检修过程中所应用的电路原理与安装软件操作方法,为日常快速修复B型彩色超声波诊断设备提供参考。Firstly, the structure and work flow of PHILPS ultrasound machine were analyzed, and the fault of electronic box of PHILPS M2540A and PHILPS HD15T color Doppler ultrasound of the failure of hardware and software are used to analysis the application of circuit principle in the process of maintenance and installation of software operation method, provides the reference for the daily fast repair of B color ultrasonic diagnostic equipment.

关 键 词:彩超机 发射电路板 视频采集卡 系统设置 