简化L波段探空仪基值测定操作方法的探讨  被引量:2

Discussion on simplified operation method of L-band radiosonde base value measurement

在线阅读下载全文

作  者:黄荟[1] 卢雪勤[1] Huang Hui;Lu Xueqin(Baise Meteorological Bureau,Baise Guangxi 53300)

机构地区:[1]百色市气象局,广西百色533000

出  处:《气象研究与应用》2018年第3期104-105,共2页Journal of Meteorological Research and Application

摘  要:通过日常的高空气象探测工作实践积累和测试分析研判,在对L波段GTS1数字探空仪进行基值测定时,基测箱上的"T/To"的开关一直向上打开,可用基测箱内的温度T(称为温度基测值)代替低湿瓶内的温度To,这样操作方法简便可行,既消除了发生误操作,又提高了工作效率,使工作进一步简化。Through the accumulation of daily high-altitude meteorological exploration work and test analysis, when the L- band GTSI digital radiosonde is measured, the "T/To" switch on the baseline box opens up all the time, and the temperature "T" in the baseline box (which called the temperature baseline value) can replace the temperature "To" in the low-humidity bottle. This operation method is simple and convenient. It is feasible to eliminate misoperation and improve work efficiency, making the work be simplified further.

关 键 词:探空仪 基值测定 湿度 To T 

分 类 号:P412.23[天文地球—大气科学及气象学]

 