不同温度及保存时间对糖化血红蛋白检测结果的影响  被引量:2

Effect of specimen disposal on detection of glycosylated hemoglobin by OLIMPO specific protein analyzer

在线阅读下载全文

作  者:黄艳萍[1] 祝峰[1] HUANG Yanping;ZHU Feng(Department of Laboratory Medicine,Jiangsu Provincial Corps Hospital of Chinese People's Armed Police Force,Yangzhou 225002,China)

机构地区:[1]武警江苏总队医院检验科,扬州225003

出  处:《武警医学》2018年第9期868-870,共3页Medical Journal of the Chinese People's Armed Police Force

摘  要:目的探讨标本在不同温度及不同保存时间的条件下,对糖化血红蛋白(Hb A1c)检测结果的影响。方法取10例患者抗凝新鲜全血标本进行预处理(溶血处理),在室温下分装成6组,每组10个标本。A组立即检测,B、C、D、E、F组分别在分装后1、2、3、4、5 h检测。取5例全血标本分别充分混匀后,分4组保存,每组5个标本。A1组在室温(18~22℃)、B1组冷藏(2~8℃)、C1组冷冻(-18^-20℃)、D1组低温冷冻(-80℃)条件下保存不同时间后,分别用OLIMPO进行检测。以低温冷冻组测定结果为标准,每个标本的Hb A1c测定结果都在低温冷冻组结果的x珋±3s范围内为可接受结果。结果溶血后不同时间Hb A1c检测结果:A组(7.98±1.61)%;B组(7.97±1.61)%;C组(7.95±1.60)%;D组(7.94±1.61)%;E组(7.77±1.61)%;F组(7.46±1.57)%。与A组比较,F组差异有统计学意义(t=1.57,P<0.05)。要使OLIMPO检测Hb A1c结果保持稳定,则标本的保存时间分别为:室温2 d,冷藏15 d,冷冻3 d;低温冷冻半年以上。结论 OLIMPO检测Hb A1c时,不同处理条件下,Hb A1c结果的稳定性差异较大。样本溶血处理后4 h内完成检测,结果较准确;如果全血标本需要保存较长的时间,则在2~8℃冷藏和-80℃条件下保存效果较好。Objective To explore the effect of sample hemolysis time and preservation conditions on the detection of glycosylated hemoglobin( Hb A1 c) by OLIMPO. Methods 1. Anticoagulant samples of fresh whole blood were taken from ten patients to be preprocessed( hemolyzed). Sixty samples of 150 μl each were collected and equally divided into 6 groups at room temperature. Group A was immediately detected,while groups B,C,D,E and F were detected 1,2,3,4 and 5 hours later,respectively. 2. After5 samples of whole blood were mixed properly,they were divided into 4 groups with 5 samples in each. Group A1 was stored at room temperature( 18-22 ℃),group B1 was refrigerated( 2-8 ℃),group C1 cryopreserved(-18--20 ℃),and group D1 was cryopreserved at a low temperature( below-80 ℃). The results of D1 were taken as the standard to detect the quality of each group with OLIMPO. The Hb A1 c results of each specimen were regarded as acceptable if they were within the range of ± 3 compared with the results of D1. Results 1. Hb A1 c results at different time points after hemolysis were( 7. 98 ± 1. 61) % in group A,( 7. 97 ± 1. 61) % in group B,( 7. 95 ± 1. 60) % in group C,( 7. 94 ± 1. 61) % in group D,( 7. 77 ± 1. 61) % in group E and( 7. 46 ± 1. 57) % in group F. There was statistically significant difference between group A and group F( t = 1. 57,P〈0. 05). 2. In order to keep Hb A1 c results stable during OLIMPO test,the storage time of samples was 2 D at room temperature,15 d for cold storage,3 D for cryopreservation and more than half a year for cryopreservation at a low temperature. Conclusions When Hb A1 c is detected with OLIMPO,the stability varies considerably under different processing conditions. The results are relatively accurate within 4 hours after sample hemolysis. If the whole blood sample specimen needs to be stored for a longer time( for scientific research or re examination),better preservation can be achieved at 2-8 ℃ and-80 ℃.

关 键 词:溶血时间 保存温度 保存时间 HBA1C OLIMPO特定蛋白分析仪 

分 类 号:R446.1[医药卫生—诊断学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象