作 者:洪扁[1] 潘征宇[1] 任学弟[1] 胡安伦[1] 陈光[1] Hong Bian;Pan Zhengyu;Ren Xuedi;Hu Anlun;Chen Guang(Shanghai Institute of Measurement and Testing Technology)
机构地区:[1]上海市计量测试技术研究院
出 处:《上海计量测试》2016年第4期30-32,共3页Shanghai Measurement and Testing
基 金:上海市质量技术监督局公益项目(2013-09)
摘 要:通过可控间隙活塞式压力计与自由形变活塞式压力计的比对实验,复现了高压自由形变活塞式压力计的形变系数。实验结果和理论计算存在较大的差异,自由形变活塞式压力计在高压力下有效面积的准确度受形变系数的影响非常大,因此对于其形变系数的不确定度需重新评估和进一步研究。The paper recurrences the high pressure balance distortioncoefficient by comparison experiments between controlled-clearancepiston gauge and free deformation pressure balance. There exists greatdifference between the results from experiment and calculation. Underhigh pressure the accuracy of effective area of free deformation pressurebalance is affected by its distortion coefficient very much,therefore theuncertainty of its distortion coefficient needs to be reevaluated and furtherresearched.
关 键 词:可控间隙活塞式压力计 自由形变活塞式压力计 形变系数 比对实验
分 类 号:TH823.2[机械工程—仪器科学与技术]