ABX PENTRA60全自动血细胞分析仪故障维修四例  被引量:2

Four Troubleshooting Cases of ABX PENTRA60 Automatic Blood Cell Analyzer

在线阅读下载全文

作  者:张涵[1] ZHANG Han(Department of Equipment, Maanshan Municipal Hospital of Traditional Chinese Medicine, Maanshan Anhui 243000, China)

机构地区:[1]马鞍山市中医院设备科,安徽马鞍山243000

出  处:《中国医疗设备》2017年第8期98-100,共3页China Medical Devices

摘  要:本文通过分析我院2015~2016年ABX PENTRA60全自动血细胞分析仪的4例故障,探讨维护思路和方法。研究发现,我院血细胞分析仪故障主要集中在管路,电磁阀,采样针等。采取定期保养措施可减少故障发生率,熟悉设备管路、电路图、散点图能快速找到故障问题所在,进行针对性维修,保障临床工作的正常开展。Four troubleshooting cases of ABX PENTRA60automatic blood cell analyzer in our hospital were analyzed,the maintenance thoughts and methods were discussed.It was found that most of the faults were concentrated on pipeline,magnetic valve and sampling needle through researching.Faults rate could be decreased by adopting periodic maintenance,and the reason of faults could be found quickly in case that medical engineer were familiar with pipeline diagram,circuit diagram and scatter plot,which could promote targeted maintenance and guarantee clinic working.

关 键 词:血细胞分析仪 电磁阀 采样针 本底 负压值 

分 类 号:TH772[机械工程—仪器科学与技术]

 