检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
作 者:胡潇寅[1] 康张宜[1] 虞跃凌[1] 汤亚杰[1] Hu Xiaoyin;Kang Zhangyi;Yü Yüeling;Tang Yajie(Shanghai Institute of Measurement and Testing Technology)
机构地区:[1]上海市计量测试技术研究院
出 处:《上海计量测试》2017年第4期36-38,共3页Shanghai Measurement and Testing
摘 要:利用磁栅尺与数字图像技术相结合的方式,开发了一种针对大量程引伸计的校准装置。将所设计的校准装置与Renishaw公司的XL-80激光干涉仪同时对一台大量程引伸计进行校准。实验结果表明,所设计的大量程引伸计校准装置在25~500 mm测量范围内的最大允许误差为0.09%。最后,对校准装置的不确定度进行了分析。In this paper,based on the integration of magnetic railing ruler and digital image correlation technology,a new calibration device for the wide range extensometer is developed.By calibrating a large range extensometer simultaneously with a laser interferometer(Renishaw,model:XL-80),the measuring range and the precision of the designed calibration device is verified.Results show that the maximum error in the range of25-500mm is0.09%.The uncertainty of the calibration device is analysed.
分 类 号:TH87[机械工程—仪器科学与技术]