检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
作 者:剡婷 凌海波[1] 蔡永前[1] YAN Ting;LING Haibo;CAI Yongqian(Center of Medical Equipment, People’s Hospital of the Xinjiang Uygur Autonomous Region, Urumqi Xinjiang 830001, China)
机构地区:[1]新疆维吾尔自治区人民医院医疗器械中心,新疆乌鲁木齐830001
出 处:《中国医疗设备》2017年第10期100-101,109,共3页China Medical Devices
摘 要:本文描述了飞利浦16排Brilliance CT Big Bore的常见维修故障3例,探寻故障排除方法及解决途径。文章首先分析了CT的组成部件及工作原理,然后根据故障现象逐步分析、探究故障原因,最终找到了CT故障原因,并给予修复。CT在临床使用中常见的故障有一定的普遍性,探讨最优的维修策略可以节约维修工时以及降低设备的维护成本,从而更好地保障临床工作。This paper described3common failures of Philips16rows of Brilliance CT Big Bore in daily work,and discussed itsmaintenance methods and solutions.Firstly,components and working principle of CT was analyzed.Secondly,causes of these faultswere explored according to failure phenomenon.Finally,causes of these faults were found and failures were repaired.The commonfailures of CT in clinical use have generality,exploring of the optimal maintenance strategy can save maintenance time and reduceequipment maintenance cost,so as to guaranteen the clinical work.
关 键 词:CT机架 CT工作站 机架控制单元 继电器 肿瘤治疗
分 类 号:TH774[机械工程—仪器科学与技术]