检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
作 者:王蒙 孙欣[3] 魏本征[1,2] WANG Meng;SUN Xin;WEI Benzheng(College of Science and Technology,Shandong University of Traditional Chinese Medicine, Jinan Shandong 250355, China;Computational Medicine Lab, Shandong University of Traditional Chinese Medicine, Jinan Shandong 250355, China;Department of Medical Engineering, Shandong Provincial Hospital Affiliated to Shandong University,Jinan Shandong 250021, China)
机构地区:[1]山东中医药大学理工学院,山东济南250355 [2]山东中医药大学计算医学实验室,山东济南250355 [3]山东大学附属省立医院医学工程部,山东济南250021
出 处:《中国医疗设备》2018年第1期39-42,共4页China Medical Devices
基 金:山东省自然科学基金项目(ZR2015FM010);山东高校科技计划项目(J15LN20);山东省中医药科技发展计划项目(2015-026);山东省医药卫生科技发展计划项目(2016WS0577)
摘 要:"O曲线"的检测对于血压模拟仪的质量控制工作以及无创血压的量值溯源均具有重要意义。在研究分析血压模拟仪输出振荡波的原理与规律的基础上,本文提出了"O曲线"的定义及其测量方法,借鉴虚拟仪器思想设计开发了血压模拟仪"O曲线"检测设备,使用该设备对Fluke公司的BP Pump 2血压模拟仪的"O曲线"进行实际测量,并以此为基础实现了对该血压模拟仪输出"O曲线"的重复性和准确性的定量评估。本研究及其开发的检测设备为血压模拟仪的质控和计量工作提供了可行的技术手段,为制定血压模拟仪相关技术标准提供了支持,具有较大的应用推广价值。The detection of“O curve”has great value to the quality control of blood pressure simulator and the traceability of noninvasive blood pressure.Based on the analysis of the principle and rules of output oscillating wave of blood pressure simulator,the novel measurement method of“O curve”was proposed in this paper,and the“O curve”detection equipment of blood pressure simulator was also designed as following the virtual instrument design technology.The experimental results showed that the developed equipment could achieve the accurate quantitative assessment of the repeatability and accuracy of the output“O curve”on the testing Fluke’s BP Pump2blood pressure simulator.This study provides a feasible technical means on the test equipment for blood pressure simulator quality control and measurement,so the research products have great application promotion value and the related further study also has reference value for the technical standards of blood pressure simulator.
关 键 词:“O曲线” 血压模拟仪 电子血压计 振荡波 质量控制
分 类 号:TH77[机械工程—仪器科学与技术]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.40