基于回归分析的尿流率计校准装置排水量模型研究  被引量:1

Research on displacement model of urine flow rate meter calibration device based on regression analysis

在线阅读下载全文

作  者:姬军[1] 张金萍[2] 张延[3] 王莉新[1] 高佳硕[1] 肖宏[1] JI Jun;ZHANG Jin-ping;ZHANG Yan;WANG Li-xin;GAO Jia-shuo;XIAO Hong(Medical Engineering Department,the 305th Hospital of the PLA,Beijing 100017,China;Medical Administration Department,the 305th Hospital of the PLA,Beijing 100017,China;Cardiothoracic Surgery Department,the 305th Hospital of the PLA,Beijing 100017,China)

机构地区:[1]解放军305医院医学工程科,北京100017 [2]解放军305医院医务部,北京100017 [3]解放军305医院心胸外科,北京100017

出  处:《医疗卫生装备》2018年第1期16-18,36,共4页Chinese Medical Equipment Journal

基  金:军事医学技术基础科研项目(军装计2016[599]);军事医学计量专项(2012-JL1-016)

摘  要:目的:寻找尿流率计校准装置齿轮泵理论排水量与实际排水量之间的相关关系,构建实际排水量与理论排水量之间的模型,以通过齿轮泵理论排水量得出实际排水量。方法:利用B2000S电子天平称量尿流率计校准装置各流率下标准水流的质量,画出理论排水量与实际排水量之间的散点图,寻找最佳拟合曲线,利用MATLAB的Curve Fitting Tool进行拟合,得到理论值与实际值之间的拟合关系式。结果:通过拟合关系式得出的拟合结果与实际排水量的误差最大为1.494%,最小为-0.010%,未超过尿流率计测量不超过5%的要求。结论:利用回归分析的方法构建尿流率计校准装置理论排水量与实际排水量之间的关系,拟合误差小,方法简单,易于实现。Objective To determine the relationship between the theory and actual displacement of the urine flow rate meter calibration device gear pump,and to construct a model between actual and theory displacement,in order to obtain actual displacement via theory displacement.Methods The weight of the standard water flow in each flow rate was measured by B2000S digital scales,and the scatter plot between the theory displacement and actual displacement was drawn to find the best fitted curve;then the MATLAB Curve Fitting Tool was used to obtain the best quasi relational equation.Results The error between the fitted displacement and the actual displacement of the gear bump was 1.494%in maximum and-0.010%in minimum,far less than 5%of the requirements for urinary flow meter measurement.Conclusion Based on regression analysis,the relationship between the theoretical displacement and actual displacement of the flow rate meter calibration device is established,the fitting error is small,and the method is simple and easy to achieve.

关 键 词:尿流率计校准装置 排水量 回归分析 MATLAB 拟合曲线 

分 类 号:R318.6[医药卫生—生物医学工程] TH789[医药卫生—基础医学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象