检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
作 者:杜晓波[1] 付成伟[1] 纪媛[1] 吕佳明 杨博闻 夏书涵 孙昕[1] 梁桁楠[1] 崔航[1] 龙北玉[1] DU Xiao-bo;FU Cheng-wei;JI Yuan;Lv Jia-ming;YANG Bo-wen;Xia Shu-han;SUN-Xin;LIANG Heng-nan;CUI Hang;LONG Bei-yu(Department of Physics,Jilin University,Changchun 130012,China)
出 处:《物理实验》2018年第11期17-19,23,共4页Physics Experimentation
摘 要:为解决稳态核磁共振实验中氟原子核横向弛豫时间T_2传统测量方法操作困难、测量精度差的问题,改进了实验仪器和测量方法.制作了测量线圈和电子积分器并用核磁共振方法进行标定,获得50Hz扫场信号.将此扫场信号输入到数字示波器x轴,获得了以50Hz交变磁场为横轴的共振吸收信号.用Origin软件对静态的吸收信号拟合处理,计算出T_2,误差小于3%.实验结果表明:该方法借助于Origin软件的数据处理能力,避免了传统测试方法中吸收信号抖动,测量操作更为容易.In order to simplify the measurement of relaxation time T 2 in the conventional nuclear magnetic resonance experiment,experimental instruments and measuring methods were improved.Detecting coil and integrator were made and calibrated to achieve the 50 Hz scanning signal,and the magnetic field dependent absorption signal was recorded.The absorption signal as well as the interference signal was fitted and calculated in Origin software,and the T 2 was obtained.The measurement result verified that this method was effective and precise.