雅培ARCHITECT i2000SR化学发光免疫分析仪报警数据分析  被引量:5

Abbott ARCHITECT i2000SR Chemiluminescence Immunoanalyzer Alarm Data Analysis

在线阅读下载全文

作  者:朱俊[1] 陈影[1] 邢昕[1] 王庆保[1] 沙泉[2] ZHU Jun;CHEN Ying;XING Xin;WANG Qingbao;SHA Quan(Physical Examination Center,The First Affiliated Hospital of Anhui University of Traditional Chinese Medicine,Hefei Anhui 230031,China;Anhui Medical University,Hefei Anhui 230032,China)

机构地区:[1]安徽中医药大学第一附属医院体检中心,安徽合肥230031 [2]安徽医科大学,安徽合肥230032

出  处:《中国医疗设备》2019年第1期174-176,共3页China Medical Devices

基  金:国家自然科学基金(81273245)

摘  要:目的保证雅培ARCHITECT i2000SR化学发光免疫分析仪日常检测顺利进行。方法对该分析仪的报警数据进行分组分析,对常见的重要故障进行原因分析、提出处理措施,并在设备修复后选择合适的性能验证方式。结果 A~E组为故障程度较低级别,报警发生率为85.49%(10259/12000);F~J组为较高级别,报警发生率为14.51%(1741/12000),经比较χ2=4020.05,P<0.01,差异有明显统计学意义。其中A组和B组报警次数最多,总共达到82.13%(9856/12000);G组和E组最少,报警发生率为0.18%(21/12000)。因吸样错误引起的报警代码有3350、2256和0842,故障发生率最高,为27.12%(3254/12000),明显高于其它报警代码。结论及时了解和处理仪器设备发出的报警信息,进行原因分析并采取处理措施,能够保证仪器顺利运行,创造良好的经济效益和社会效益。Objective To ensure the daily detection of Abbott ARCHITECT i2000SR chemiluminescence immunoanalyzer carried out smoothly.Methods The alarm data of the analyzer were grouped for analysis,causes of common important faults were analyzed,treatment measures were proposed,and appropriate performance verification methods were selected after equipment repair.Results Groups A^E had a low degree of failure,with an alarm rate of 85.49%(10259/12000).The F^J groups had a relatively high level,with an alarm rate of 14.51%(1741/12000).After comparison(χ^2=4020.05,P<0.01),the difference was statistically significant.Group A and group B had the highest number of alarms,reaching 82.13%(9856/12000)in total.Group G and group E had the lowest alarm rate of 0.18%(21/12000).The alarm codes caused by sampling error were 3350,2256 and 0842,with the highest failure rate of 27.12%(3254/12000),significantly higher than other alarm codes.Conclusion Timely understanding and processing of alarm information issued by instruments and equipment,cause analysis and processing measures can ensure the smooth operation of instruments and create good economic and social benefits.

关 键 词:雅培ARCHITECT i2000SR 化学发光免疫分析仪 报警数据分析 故障处理 

分 类 号:TH776[机械工程—仪器科学与技术] R197.39[机械工程—精密仪器及机械]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象