棱镜摄谱仪中紫光谱成像焦距变长的原因  

Investigation of longer focus length of violet spectral imaging in prism spectrograph

在线阅读下载全文

作  者:张权[1] 赵伟[1] 王鹤[1] 朱玲[1] 王中平[1] 王晨[1] 郑虹[1] ZHANG Quan;ZHAO Wei;WANG He;ZHU Ling;WANG Zhong-ping;WANG Chen;ZHENG Hong(Department of Astronomy and Applied Physics,University of Science and Technology of China,Hefei 230026,China)

机构地区:[1]中国科学技术大学物理实验教学中心,安徽合肥230026

出  处:《物理实验》2018年第B12期22-25,共4页Physics Experimentation

摘  要:在使用WPL小型棱镜摄谱仪记录光谱时,紫光成像像距比红光成像像距长.本文使用光学计算软件ZEMAX对小型棱镜摄谱仪光路中各元件的光学参量进行了详细计算,利用系统光具组计算公式,从数值上准确地给出了小型棱镜摄谱仪中紫光成像像距长于红光成像像距的计算结果和详细分析.While recording spectra by using the WPL miniature prism spectrograph,the image distance of violet light is longer than that of red light.In this paper,ZEMAX,an optical calculation software,was used to calculate the optical parameters of each component in the optical path of the prism spectrograph.Calculation results and detailed analysis of longer violet image distance in the prism spectrograph were also given numerically and accurately by using the calculation formula of optical system.

关 键 词:棱镜摄谱仪 紫光谱 像距 

分 类 号:O435[机械工程—光学工程]

 