PB840呼吸机流量传感器工作原理及故障修复方法探索  被引量:7

Working Principle and Troubleshooting Method of PB840 Ventilator Flow Sensor

在线阅读下载全文

作  者:李海丰[1] 郑沁春[1] LI Haifeng;ZHENG Qinchun(Department of Equipment,Fujian Provincial Hospital,Fuzhou Fujian 350001,China)

机构地区:[1]福建省立医院设备处,福建福州350001

出  处:《中国医疗设备》2019年第3期174-176,共3页China Medical Devices

摘  要:目的对呼吸机质量控制检测中发现的难以修复的流量传感器问题进行尝试性修复。方法通过酒精和酶洗的方法,修复流量传感器。结果经质量控制检测,经过修复的流量传感器,符合标准。结论酒精和酶洗的方法可快速有效地修复流量传感器,可提高维护效率,降低维护成本。Objective To repair the flow sensor which was difficult to repair in the ventilator quality control test. Methods The flow sensor was repaired by alcohol and enzyme washing. Results The flow sensor repaired by alcohol and enzyme washing was qualified by quality control. Conclusion Alcohol and enzyme washing methods can quickly and effectively repair flow sensors, which can improve maintenance efficiency and reduce maintenance costs.

关 键 词:呼吸机 流量传感器 故障检测 修复方法 

分 类 号:R197.39[医药卫生—卫生事业管理] TH789[医药卫生—公共卫生与预防医学]

 