640层螺旋CT低剂量扫描技术结合低浓度对比剂在冠状动脉支架CTA中的应用  被引量:24

The application value of 640-slice spiral CT with low radiation dose scanning technique and low dose contrast agent in coronary stent CTA

在线阅读下载全文

作  者:荣梅[1] 李彩辉[2] 闫建华 赵启利[1] 李兆祥[1] RONG Mei;LI Caihui;YAN Jianhua(Department of Radiodiagnosis,People’s Hospital of Langfang City,Hebei,Langfang 065000,China)

机构地区:[1]河北省廊坊市人民医院放射诊断科,065000 [2]廊坊卫生职业学院临床与预防医学系 [3]河北省廊坊市人民医院供应室,065000

出  处:《河北医药》2019年第5期645-649,654,共6页Hebei Medical Journal

基  金:廊坊市科学技术研究与发展计划项目(编号:2014013008C)

摘  要:目的探讨640层螺旋CT低剂量扫描技术结合低浓度对比剂在冠状动脉支架CTA中的应用价值。方法前瞻性连续入选2014年10月至2015年10月冠脉支架CTA随访门诊患者60例,随机分为A、B 2组,每组30例。A组应用等渗低浓度对比剂威视派克(270 mgI/ml)低剂量(100 Kvp)扫描;B组用高浓度对比剂欧乃派克(350 mgI/ml),常规剂量(120 kVp)扫描。患者在冠状动脉支架CTA(coronary stent CTA,CSCTA)检查后1~2周内均行冠状动脉造影检查(coronary angiography,CAG)。比较2组主动脉根部CT值、噪声值(SD)、图像信噪比(SNR)、对比噪声比(CNR)、图像质量评分、CT容积剂量指数(CTDIvoI)、剂量长度乘积(DLP)、有效辐射剂量(ED)及碘摄入量;以CAG检查结果为金标准,比较2组CSCTA与支架CAG检查结果,评价冠脉支架狭窄的检出率差异。结果共检出81枚冠脉支架,其中78枚支架可评价(A组43枚,B组35枚)。2组的SD、SNR、CNR和图像质量评分差异无统计学意义(P>0.05);2组的CTDIvol、DLP、ED以及碘摄入量差异均有统计学意义(P<0.05),A组均明显低于B组,其中A组的有效辐射剂量(ED)较B组下降约61.72%。2组均有较高的冠脉支架再狭窄检出率,2组检出率之间差异无统计学意义(P>0.05)。结论在冠状动脉支架术后CTA随访中,在保证图像质量和冠脉支架再狭窄检出率的前提下,低管电压结合低浓度对比剂扫描能明显降低辐射剂量,并减少对比剂有效碘用量,因此低剂量低浓度对比剂扫描将成为今后冠状动脉支架CTA随访的必然趋势。Objective To investigate the application value of 640-slice spiral CT with low radiation dose scanning technique and low dose contrast agent in coronary stent CTA(CSCTA).Methods A total of 60 patients who underwent coronary stent CTA in our hospital from October 2014 to October 2015 were randomly divided into group A(n=30)and group B(n=30).The patients in goup A were underwent low radiation dose scanning with isoosmia low dose contrast agent-visipaque(270mgI/ml,100kVp),and the patients in group B underwent conventional dose scanning with high dose contrast agent-omnipaque(350mgI/ml,120kVp).The patients in both groups were examined by the coronary angiography(CAG)within 1~2 weeks after CSCTA.The aorta root’s CT value,noise value(SD),image signal-to-noise ratio(SNR),comparison noise ratio(CNR),image quality score,volume CT dose index(CTDIvol),product of dose length(DLP),the effective radiation dose(ED),and the iodine intaking dose as well as the examination results of CSCTA and CAG were observed and compared between the two grousp.Results A total of 81 coronary stents were detected,in which 78 pieces were evaluable(43 pieces in group A,35 pieces in group B).There were no significant differences in aorta root’s CT value,noise,SNR,CNR and scores of the image quality between group A and group B(P>0.05).However there were significant differences in CTDIvol,DLP,ED and iodine intaking dose between group A and group B,which in group A were significantly lower than those in group B(P<0.05),and the ED in group A was decreased by 61.72%,as compared with that in group B.The detection rate of in-stent restenosis(ISR)was higher in both groups,but there was no significant difference in the detction rate(ISR)between the two groups(P>0.05).Conclusion In follow up after CSCTA,on the premise of guaranteeing the quality of image and the detection rate of ISR,to use lower tube voltage scanning combined with low-dose contrast agent can significantly decrease the radiation dose,and reduce the effective iodine contrast agent dosage.T

关 键 词:640层螺旋CT 低辐射剂量 低浓度对比剂 冠状动脉支架CTA 

分 类 号:R814.42[医药卫生—影像医学与核医学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象