GE Senographe DS乳腺机维修案例分析  被引量:2

Maintenance Case Analysis of GE Senographe DS Mammography Machine

在线阅读下载全文

作  者:熊贵富 XIONG Guifu(Department of Medical Equipment, Yongchuan Hospital of Chongqing Medical University, Chongqing 402160, China)

机构地区:[1]重庆医科大学附属永川医院设备科,重庆402160

出  处:《中国医疗设备》2019年第4期171-173,共3页China Medical Devices

摘  要:本文通过分析GE Senographe DS乳腺机不能曝光及机架C形臂不能旋转的故障,为医学工程师们维修类似故障提供参考和借鉴。对于这两种故障,通过查看错误日志,翻阅维修手册,找到问题所在,并及时排除了故障,维修后设备经过一周的运行,各项指标正常。以上两种故障,维修方法可行和有效,且维修成本低。In this paper, the failure of no X-ray and immovable of C-arm of GE Senographe DS mammography machine was analyzed, so as to provide reference for medical engineers to repair similar failure. According to two kinds of failure phenomenon, the faults were found and solved by reviewing the system event log and maintenance manual. After a week of running, all indexes were normal. The respective method was feasible and effective for both malfunction above, and could reduce the maintenance cost.

关 键 词:乳腺机 高压油箱 球管灯丝 机架C臂 电位器 

分 类 号:R197.39[医药卫生—卫生事业管理]

 