荧光法快速测定水中粪大肠菌群  被引量:5

Rapid Determination of Fecal Coliforms in Water by Fluorescence Method

在线阅读下载全文

作  者:由希华[1] 徐标[1] 李恒庆[1] 宗雪梅[1] YOU Xihua;XU Biao;LI Hengqing;ZONG Xuemei(Environmental Monitoring Centre of Shandong Province ,Jinan 250101 ,China)

机构地区:[1]山东省环境监测中心站,山东济南250101

出  处:《中国环境监测》2019年第2期99-103,共5页Environmental Monitoring in China

摘  要:为研究荧光法测定水中粪大肠菌群的适用性,在6家实验室使用荧光法对粪大肠菌群的标准菌株和实际水样进行测定,并与多管发酵法进行比对。结果表明:6家实验室对标准菌株的监测结果均在误差范围内,实验室内相对误差为-2. 75%~1. 71%,准确度较高;对3个不同浓度实际水样进行精密度测定,实验室内相对标准偏差分别为1. 30%~3. 93%、1. 94%~4. 72%、1. 88%~4. 54%,实验室间的相对标准偏差分别为1. 14%、1. 59%、1. 72%,精密度较高;用荧光法与多管发酵法同时测定实际水样,2种方法测试结果线性相关性良好且变化趋势相同,对测试数据进行t检验,2种方法测试结果无显著差异。In order to verify the suitability of fluorescence method for the determination of fecal coliforms in water, we tested the standard strains of fecal coliforms in six laboratories using the Method, and did a comparison with multi-tube fermentation method. The results were all with the acceptable range. The relative error in the same laboratory was -2.75%-1.71%, which showed a relatively high accuracy. When testing the regular water samples with three different concentrations, the Method provided the relative standard deviations for each laboratories as follows:1.30%-3.93%, 1.94%-4.72%, 1.88%-4.54%.The relative standard deviations for the six laboratories were 1.14%, 1.59%, 1.72%, which provided a high precision as well. The samples from surface water or sewage samples were measured by the Method and the multi-tube method, the two methods gave a good linear correlation and a same trend. When evaluating the results using the t test, it showed that there was no significant difference between the two methods. The method is simple, fast and accurate, and can be used as a standard method for evaluating the microorganism in water.

关 键 词:粪大肠菌群 荧光法 多管发酵法 

分 类 号:X832.02[环境科学与工程—环境工程]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象