人工肛门括约肌无线供能系统设计与优化  被引量:3

Design and Optimization of Wireless Energy Supply System for Artificial Anal Sphincter

在线阅读下载全文

作  者:吴昌建 姜志华 颜国正[1] 周泽润 王志武[1] 赵凯[1] 韩玎 姜萍萍[1] WU Changjian;JIANG Zhihua;YAN Guozheng;ZHOU Zerun;WANG Zhiwu;HAN Ding;JIANG Pingping(School of Electronic Information and Electrical Engineering,Shanghai Jiao Tong University,Shanghai 200240,China;Shanghai Institute of Measurement and Testing Technology,Shanghai 201203,China)

机构地区:[1]上海交通大学电子信息与电气工程学院,上海200240 [2]上海市计量测试研究院,上海201203

出  处:《上海交通大学学报》2019年第9期1045-1050,共6页Journal of Shanghai Jiaotong University

基  金:国家自然科学基金(61673271,81601631);上海市科研计划(15441903100);中国博士后基金(2016M601587)资助项目

摘  要:基于电磁耦合原理,针对人造肛门括约肌系统,设计了一套无线供能系统.该系统主要包括无线能量发射端、接收端、整流模块、稳压模块和充电模块.研究了发射端的发射频率(40~120 kHz)对传输功率的影响,进而优化了稳压模块,并比较了锂电池充电效果.结果表明:在一定发射频率范围内,无线供能传输效率随着发射频率的增大而升高;发射频率为120 kHz时,传输效率可以达到57.47%,接收功率为1.12 W;锂电池能快速充电,为人造肛门括约肌系统提供稳定能量保障.Based on the principle of electromagnetic coupling,a set of wireless energy supply system was designed for the artificial anal sphincter system.The system contains five parts:wireless energy transmitter,wireless energy receiver,rectifier module,regulator module and charging module.This work focuses on the effect of transmitter launch angle frequency(40-120 kHz)on the transmission power.We compare the charging effect and optimize the regulator module to meet the executing agency operating power requirements.The experiment results show that the transmission efficiency improves with the increase of the transmission frequency in a certain frequency range.When the transmission frequency is 120 kHz,the transmission efficiency can reach 57.47%,the receiving power is 1.12 W,and the lithium battery can be quickly charged,providing stable energy for the artificial anal sphincter system.

关 键 词:人工肛门括约肌 无线供能 稳压模块 充电模块 

分 类 号:R314TM15[医药卫生—基础医学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象