牙科X射线诊断的受检者剂量调查研究  被引量:3

A survey of examinee doses in dental X-ray examinations

在线阅读下载全文

作  者:冯泽臣[1] 张志彬 王宏芳[1] 白斌[1] 赵思京 石红梅[2] 赵红枫[3] 娄云[1] FENG Zechen;ZHANG Zhibin;WANG Hongfang;BAI Bin;ZHAO Sijing;SHI Hongmei;ZHAO Hongfeng;LOU Yun(Beijing Center for Disease Control and Prevention,Beijing Preventive Medicine Research Center,Beijing 100013 China;Beijing Dongcheng District Center for Disease Prevention and Control;Beijing Chaoyang District Center for Disease Prevention and Control)

机构地区:[1]北京市疾病预防控制中心北京市预防医学研究中心,北京100013 [2]北京市东城区疾病预防控制中心 [3]北京市朝阳区疾病预防控制中心

出  处:《中国辐射卫生》2019年第3期271-275,共5页Chinese Journal of Radiological Health

基  金:北京市疾病预防控制中心,北京市预防医学研究中心科研培育专项(2018-BJYJ-08)

摘  要:目的调查牙科X射线诊断所致受检者剂量范围,为牙科X射线检查辐射剂量最优化提供参考,降低牙科X射线检查的辐射风险。方法使用多功能X射线测量仪Raysafe X2和PTW Diamentor CX型空气比释动能面积乘积仪在北京市6家医疗机构,对口内牙科X射线机的入射空气比释动能和空气比释动能面积乘积进行了调查,对口外牙科X射线机的空气比释动能面积乘积进行了调查。结果共调查了5台口内牙科X射线机和6台口外牙科X射线机,口内牙科X射线机的入射空气比释动能范围为0.57 mGy^2.95 mGy,KAP范围为13.2 mGy·cm^2~70.8 mGy·cm^2。口外牙科X射线机全景的空气比释动能面积乘积范围为54.2 mGycm^2~139.4 mGy·cm^2,头颅正侧位的空气比释动能面积乘积范围为24.8 mGycm^2~83.6 mGy·cm^2,CBCT的空气比释动能面积乘积范围为714.7 mGy·cm^2~3 791.5 mGy·cm^2。结论牙科X射线应根据诊断目的合理选择辐射风险低的检查类型,对不同类型检查根据所要达到的图像质量选择合适的照射条件,通过患者剂量调查建立诊断参考水平是防护最优化的重要手段。Objective To investigate the dose range of examinees in dental X-ray examinations,and provide advice for optimizing dose to reduce the radiation risk of dental X-ray examination.Methods The Raysafe X2 and PTW Diamentor CX were used to investigate the incident air kerma and air kerma area product(KAP)of the intraoral dental X-ray machine.The PTW Diamentor CX were used to investigate air kerma area product(KAP)of the extraoral dental X-ray machine.The measurements were done at 6 medical institutions in Beijing.Results The incident air kerma of the intraoral dental X-ray examinationsranged from 0.57 mGy to 2.95 mGy,and the KAP ranged from 13.2 mGy·cm^2 to 70.8 mGy·cm^2.The KAP of the panoramaranged from 54.2 mGy·cm^2 to 139.4 mGy·cm^2,and the KAP of the cephlometricsranged 24.8 mGy·cm^2 to 83.6 mGy·cm^2,and the KAP of the CBCT ranged 714.7 mGy·cm^2 to 3 791.5 mGy·cm^2.Conclusion According to the purpose of diagnosis,the type of examination with low radiation risk should be selected.The appropriate parameters should be selected according to the required image quality.Establishing the diagnostic reference level through thedose investigation is an important means of protection optimization.

关 键 词:牙科X射线机 入射空气比释动能 空气比释动能面积乘积 

分 类 号:R144.1[医药卫生—公共卫生与预防医学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象