Impact of corneal parameters on intraocular pressure measurements in different tonometry methods  被引量:6

Impact of corneal parameters on intraocular pressure measurements in different tonometry methods

在线阅读下载全文

作  者:Aleksandra Zakrzewska Marta P.Wi?cek Anna Machalińska 

机构地区:[1]First Department of Ophthalmology, Pomeranian Medical University

出  处:《International Journal of Ophthalmology(English edition)》2019年第12期1853-1858,共6页国际眼科杂志(英文版)

摘  要:AIM: To evaluate the impact of central corneal thickness(CCT) and corneal curvature on intraocular pressure(IOP) measurements performed by three different tonometers. METHODS: IOP in 132 healthy eyes of 66 participants was measured using three different tonometry techniques: Goldmann applanation tonometer(GAT), Pascal dynamic contour tonometer(DCT), and ICare rebound tonometer(RT). CCT and corneal curvature were assessed. RESULTS: In healthy eyes, DCT presents significantly higher values of IOP than GAT(17.34±3.69 and 15.27±4.06 mm Hg, P<0.0001). RT measurements are significantly lower than GAT(13.56±4.33 mm Hg, P<0.0001). Compared with GAT, DCT presented on average 2.51 mm Hg higher values in eyes with CCT<600 μm and 0.99 mm Hg higher results in eyes with CCT≥600 μm. The RT results were lower on average by 1.61 and 1.95 mm Hg than those obtained by GAT, respectively. Positive correlations between CCT in eyes with CCT<600 μm were detected for all IOP measurement techniques, whereas a similar relationship was not observed in eyes with thicker corneas. A correlation between IOP values and keratometry in the group with CCT<600 μm was not detected with any of the tonometry methods. In thicker corneas, a positive correlation was found for GAT and mean keratometry values(R=0.369, P=0.005). CONCLUSION: The same method should always be chosen for routine IOP control, and measurements obtained by different methods cannot be compared. All analysed tonometry methods are dependent on CCT;thus, CCT should be taken into consideration for both diagnostics and monitoring.● AIM: To evaluate the impact of central corneal thickness(CCT) and corneal curvature on intraocular pressure(IOP) measurements performed by three different tonometers.● METHODS: IOP in 132 healthy eyes of 66 participants was measured using three different tonometry techniques: Goldmann applanation tonometer(GAT), Pascal dynamic contour tonometer(DCT), and ICare rebound tonometer(RT). CCT and corneal curvature were assessed.● RESULTS: In healthy eyes, DCT presents significantly higher values of IOP than GAT(17.34±3.69 and 15.27±4.06 mm Hg, P<0.0001). RT measurements are significantly lower than GAT(13.56±4.33 mm Hg, P<0.0001). Compared with GAT, DCT presented on average 2.51 mm Hg higher values in eyes with CCT<600 μm and 0.99 mm Hg higher results in eyes with CCT≥600 μm. The RT results were lower on average by 1.61 and 1.95 mm Hg than those obtained by GAT, respectively. Positive correlations between CCT in eyes with CCT<600 μm were detected for all IOP measurement techniques, whereas a similar relationship was not observed in eyes with thicker corneas. A correlation between IOP values and keratometry in the group with CCT<600 μm was not detected with any of the tonometry methods. In thicker corneas, a positive correlation was found for GAT and mean keratometry values(R=0.369, P=0.005).● CONCLUSION: The same method should always be chosen for routine IOP control, and measurements obtained by different methods cannot be compared. All analysed tonometry methods are dependent on CCT; thus, CCT should be taken into consideration for both diagnostics and monitoring.

关 键 词:intraocular pressure Goldmann applanation tonometer Pascal dynamic contour tonometer ICare rebound tonometer central corneal thickness corneal curvature healthy individuals 

分 类 号:TP3[自动化与计算机技术—计算机科学与技术]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象