医院级CT辐射剂量及诊断剂量水平研究  被引量:8

Research on hospital-level CT radiation dose and diagnostic dose level

在线阅读下载全文

作  者:王亮 彭勇[1] 刘怡[1] 彭红芬[1] WANG Liang;PENG Yong;LIU Yi;PENG Hongfen(Department of Radiology,Wuhan No.1 Hospital,Wuhan Hospital of TraditionalChinese and Western Medicine,Wuhan 430022)

机构地区:[1]武汉市第一医院武汉市中西医结合医院医学影像科

出  处:《辐射防护》2019年第6期517-521,共5页Radiation Protection

摘  要:探讨医院级各部位CT扫描辐射剂量及成人CT检查诊断剂量水平。采集2018年2月—2019年1月本院在GE Light Speed 16 CT设备上进行诊断性CT检查的全部成年人共15440例的CT检查数据,按照检查部位进行分类,统计各部位的容积剂量指数(CTDI vol)、剂量长度乘积(DLP)。使用SPSS 19统计软件分析CTDI vol、DLP的四分位数并计算有效剂量(ED)。结果表明,各部位CTDI vol的诊断剂量参考水平(DRL)值为:颅脑87.75 mGy、肺部8.09 mGy、上下腹15.82 mGy、上下腹盆腔15.79 mGy;各部位DLP的诊断剂量参考水平(DRL)值为:颅脑1053.03 mGy·cm、肺部245.19 mGy·cm、上下腹269.96 mGy·cm、上下腹盆腔835.20 mGy·cm。研究中发现肺部CTDI vol、DLP的DRL值(分别为8.09 mGy、245.19 mGy·cm)低于我国其他地区和其他国家的DRL值;60岁以下患者肺部CT检查CTDI vol、DLP的DRL值低于8.09 mGy、245.19 mGy·cm,60岁以上患者肺部CT检查CTDI vol、DLP的DRL值高于8.09 mGy、245.19 mGy·cm。通过研究可以确定医院级各部位CT检查辐射剂量及成人CT检查诊断剂量水平,能为优化CT检查扫描技术和监测辐射剂量提供基础。This work aims to study the CT scan radiation dose of the human anatomical region of hospital scale and diagnostic dose level of adult CT examination.A total of 15,440 CT scan data were collected from the anatomical regions of all adults who underwent diagnostic CT examinations on GE Light Speed 16 CT equipment from February 2018 to January 2019.The volumetric dose index(CTDI vol)and the dose length product(DLP)of the patient’s anatomical region were counted according to the anatomical region classification.The quantities of CTDI vol and DLP in the scan data were statistically analyzed using spss19 software and the effective dose(ED)was calculated.As a result,the DRL values of CTDI vol in each anatomical region were as follows:87.75 mGy for the brain,8.09 mGy for the lungs,15.82 mGy for the upper and lower abdomen,and 15.79 mGy for the upper and lower abdominal pelvis.The DRL values of DLP in each anatomical region were as follows:for the brain was 1053.03 mGy cm,for the lung was 245.19 mGy cm,for the upper and lower abdomen was 269.96 mGy cm,for the upper and lower abdominal pelvic was 835.20 mGy cm.The DRL values of CTDI vol and DLP in lung anatomical region(8.09 mGy,245.19 mGy cm)were lower than those in other regions of China and other countries.The DRL values of CTDI vol and DLP in patients under 60 year old were lower than 8.09 mGy and 245.19 mGy cm,and the DRL values of CTDI vol and DLP in CT scan of patients over 60 year old were higher than 8.09 mGy and 245.19 mGy cm.Taking all together,according to the analysis of the research results,the CT radiation dose and the adult CT examination dose level of each anatomical area of the hospital can be determined,which provides a basis for optimizing the CT scanning technique and monitoring the radiation dose.

关 键 词:辐射剂量 X线计算机断层摄影 剂量优化 诊断剂量水平 

分 类 号:R144.1[医药卫生—公共卫生与预防医学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象