作 者:王喆[1] WANG Zhe(The Eye Hospital of the Zibo City,Hebei Province,the third hospital in the city of Zibo City,Handan Hebei 056001,China)
机构地区:[1]河北省邯郸市眼科医院邯郸市第三医院
出 处:《中国医疗设备》2019年第S02期103-105,共3页China Medical Devices
摘 要:目的研究金属医疗器械采用低温过氧化氢等离子和脉动预真空压力蒸汽灭菌的效果。方法研究时段2017年1月至2019年4月,研究对象是我院接收的2000件消毒灭菌金属器械,随机数字表法划分成观察组(低温过氧化氢等离子)与对照组(脉动预真空压力蒸汽)各1000件,比较两组灭菌效果。结果观察组灭菌合格率(100.00%)高于对照组99.00%,灭菌时间较对照组长,灭菌温度较对照组低,差异具统计学意义(P<0.05)。结论低温过氧化氢等离子能提高金属医疗器械灭菌效果,便于为操作者自身安全提供可靠的保障。Objective The purpose of this study was to study the effect of low temperature hydrogen peroxide plasma and pulse prevacuum pressure steam sterilization for metal medical devices.Methods The study period was 2017.1 to 2019.4.The study object was 2000 pieces of sterilized metal device received in our hospital.The random digital surface method was divided into 1000 pieces of observation group(low temperature hydrogen peroxide plasma)and control group(pulse pre-vacuum pressure steam),and the two groups of sterilization effects were compared.Results The rate of sterilization of the observation group(100.00%)was higher than that of the control group(99.00%),the sterilization time was longer than that of the control group,and the sterilization temperature was lower than that of the control group(P<0.05).Conclusion The low temperature hydrogen peroxide plasma can improve the sterilization effect of metal medical devices and provide reliable guarantee for the operator’s own safety.
关 键 词:金属医疗器械 低温过氧化氢等离子 脉动预真空压力蒸汽
分 类 号:TH789[机械工程—仪器科学与技术]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.157