Gambro AK 96血液透析装置维修案列分析与解决  被引量:3

Maintenance Cases Analysis and Troubleshooting of Gambro AK 96 Hemodialysis Equipment

在线阅读下载全文

作  者:龚希 陶陶 GONG Xi;TAO Tao(Department of Equipment,The No.1 People’s Hospital of Pinghu,Pinghu Zhejiang 314200,China)

机构地区:[1]平湖市第一人民医院设备科

出  处:《中国医疗设备》2020年第2期178-180,184,共4页China Medical Devices

摘  要:本文通过对Gambro AK 96血液透析装置出现的两例故障进行描述分析,提供具体的解决方案,同时通过这两例维修案列探讨改进Gambro AK 96血液透析装置的日常维护。故障一主要由静脉压传感器发生故障引起,故障二主要由消毒液阀膜老化引起。我们通过查找此机型的机器报错代码,结合透析液液路图,快速准确的找到故障点,进行针对性维修。工程师需要熟悉报错代码,精通透析液液路图,快速解决故障,为医院节省维修成本。Two failures in the Gambro AK 96 hemodialysis unit were analyzed and specific solutions were provided.Based on these two maintenance cases,the measures were explored to improve the daily maintenance of the Grambro AK 96 hemodialysis unit.The first case was caused by the failure of the venous pressure module,and the second case was mainly caused by the aging of the disinfectant valve membrane.By searching the machine error code of this model and combining with the dialysate liquid circuit diagram,we can quickly and accurately find the fault point and carry out targeted maintenance.Engineers need to be familiar with error reporting code,master dialysate flow chart,quickly solve problems,so as to save maintenance costs for hospitals.

关 键 词:血透机 静脉压传感器 消毒阀膜 报错代码 透析液液路图 

分 类 号:R318.6[医药卫生—生物医学工程] TH772.1[医药卫生—基础医学]

 