除颤监护仪的质量检测探讨  被引量:9

Discussion on quality inspection of defibrillation monitor

在线阅读下载全文

作  者:张华伟[1] 田书畅 王雨荷 舒慧宇[2] ZHANG Hua-wei;TIAN Shu-chang;WANG Yu-he;SHL Hiii-yu(Department of Medical Engineering,Nanjing First Hospital Nanjing Medical University,210006,Jiangsu,China;Department of Purchasing Lopistics,Nanjing First Hospital Nanjing Medical University,210006,Jiangsu,China)

机构地区:[1]南京医科大学附属南京医院(南京市第一医院)医疗设备处,江苏南京210006 [2]南京医科大学附属南京医院(南京市第一医院)采购物流中心,江苏南京210006

出  处:《生物医学工程与临床》2020年第1期95-97,共3页Biomedical Engineering and Clinical Medicine

基  金:南京医科大学科技发展基金一般项目(NMUB2018338)。

摘  要:目的探讨除颤监护仪质量检测的方法,掌握在用除颤监护仪的使用状态。方法以某三级甲等医院40台除颤监护仪为研究对象,采用美国福禄克(FLUKE)专业的质量检测设备对除颤监护仪进行检测。结果对除颤监护仪进行质量检测,初次检测外观合格率为50%,电气安全合格率为85%,能量合格率为95%,心电合格率100%。其中外观检测中不合格的主要原因是日期时间不准,电气安全检测中不合格的主要原因是接地阻抗过高,能量检测中不合格的主要原因是释放能量偏大、充电时间过长。不合格除颤监护仪经过问题纠正后二次检测全部合格。结论除颤监护仪的质量检测关系重大,必须加强监管,定期对其主要参数进行质量检测,以确保其始终保持在待用状态。Objective To investigate quality detection method of defibrillation monitors and obtain usage status of defibrillation monitors.Methods The FLUKE professional quality testing equipment was used to test 40 defibrillation monitors in a level 3 grade A hospital.Results For pass rate of defibrillation quality detection,the initial appearance was 50%,electrical safety was 85%,energy was 95%and ECG was 100%.The main failure reasons for appearance was inaccurate of date and time,electrical safety was too high grounding impedance,energy was too big releasing energy and too long charging time.All unqualified defibrillation monitors passed the secondary detection after problems were corrected.Conclusion It is demonstrated that defibrillation monitor quality is very important,and supervision must be strengthened to periodically check the main parameters for ensuring the monitors remain ready for use at all times.

关 键 词:除颤监护仪 质量检测 电气安全 接地阻抗 

分 类 号:R319[医药卫生—基础医学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象