作 者:贺飞耀 赵忠俊[2] 任标 张超凡 付玉 段忆翔[1] HE Fei-yao;ZHAO Zhong-jun;REN Biao;ZHANG Chao-fan;FU Yu;DUAN Yi-xiang(College of Life Sciences,Sichuan University,Chengdu 610064,China;School of Chemical Engineering,Sichuan University,Chengdu 610064,China;School of Mechanical Engineering,Sichuan University,Chengdu 610064,China)
机构地区:[1]四川大学生命科学学院,四川成都610064 [2]四川大学化学工程学院,四川成都610064 [3]四川大学机械工程学院,四川成都610064
出 处:《质谱学报》2020年第2期110-117,I0003,共9页Journal of Chinese Mass Spectrometry Society
基 金:国家自然科学基金青年基金项目(21605107);中央高校基本科研经费(SCU2015c003)资助。
摘 要:离子传输系统是质谱仪的重要组成部分,主要作用是将离子高效率地传输到质量分析器。本工作研制了一种用于质子转移反应飞行时间质谱(PTR-TOF MS)系统的射频四极杆离子导向装置,四极杆长80 mm,杆半径2.6 mm,内切圆半径2.25 mm,该装置可针对性地实现低质荷比挥发性有机化合物(VOCs)离子的聚焦传输。利用SIMION 8.1离子光学模拟平台对装置的运行环境进行仿真,然后在自行搭建的测试平台上对装置的工作条件,如气压、频率和电压幅值进行测试。结果表明,仿真和测试结果具有较好的一致性,装置的工作气压范围较宽,在0.2~0.3 Pa时的传输效率最高;当频率为3~4 MHz,电压幅值(V p-p)为500 V左右时,对丙酮、甲苯等低质荷比VOCs(<m/z 100)的传输效率接近76%,且离子束直径≤0.7 mm。该装置结构简单、成本低、传输效率高,具有潜在的实用价值,有望应用于PTR-TOF MS系统。The ion guide system is an important part of the mass spectrometer, which main function is to efficiently transfer ions to the mass analyzer. In this work, a radio frequency quadrupole ion guide device for proton transfer reaction time-of-flight mass spectrometry(PTR-TOF MS) was developed and the quadrupole length was 80 mm, the rod radius was 2.6 mm, and the inscribed circle radius was 2.25 mm. This device realized focusing and transmission of volatile organic compounds(VOCs) with low mass-to-charge ratio. The movement of ions in the RF-only quadrupole device was simulated by SIMION 8.1 ion optical simulation software. The working conditions of the device, such as pressure, frequency and voltage amplitude, were tested on the home-built test platform. The results showed that the simulation and test had good consistency. The working pressure, frequency and amplitude of voltage were optimized to be 0.2-0.3 Pa, 3-4 MHz and about 500 V, respectively. The transmission efficiency of low mass-to-charge ratio VOCs(less than m/z 100) ions such as acetone and toluene was close to 76% with ion beam diameter no more than 0.7 mm. The device has the advantages of simple structure, low cost and high transmission efficiency, and which will be expected to be used in the PTR-TOF MS system.
关 键 词:离子传输 射频四极杆 质子转移反应质谱(PTR-MS) 飞行时间质谱(TOF MS)
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.229