作 者:王美娇[1] 周舜 岳海振[1] 于松茂 吴昊[1] 王海洋 张艺宝[1] WANG Meijiao;ZHOU Shun;YUE Haizhen;YU Songmao;WU Hao;WANG Haiyang;ZHANG Yibao(Key Laboratory of Carcinogenesis and Translational Research(Ministry of Education/Beijing)/Department of Radiotherapy,Peking University Cancer Hospital&Institute,Beijing 100142,China)
机构地区:[1]北京大学肿瘤医院暨北京市肿瘤防治研究所放疗科/恶性肿瘤发病机制及转化研究教育部重点实验室,北京100142
出 处:《中国医学物理学杂志》2020年第3期270-272,共3页Chinese Journal of Medical Physics
基 金:首都卫生发展科研专项(首发2018-4-1027);北京市自然科学基金(7172048,1184014);教育部科技发展中心产学研创新基金“-智融兴教”基金(2018A01019);国家自然科学基金(11505012);四川省科技计划资助项目(2018HH0099);北京市属医院科研培育计划项目(PX2019042,PX2016060);北京市医院管理局“青苗”计划专项经费(QML20171104)。
摘 要:目的:比较双面网和单面网两种固定方式对头部立体定向放射治疗病例摆位精度的影响。方法:选取40例头部立体定向放射治疗患者,分为对照组(传统单面网固定)和试验组(新型双面网固定)。获取患者治疗前锥形束CT图像,记录按照骨性标志配准误差并进行统计学分析。结果:两组摆位误差在左右(Lat)、进出(Lng)、升降(Vrt)、俯仰(Pitch)、翻转(Roll)、旋转(Rtn)各方向平均数±标准差表示为:对照组(0.26±0.15)cm、(0.13±0.10)cm、(0.19±0.13)cm、(1.40±0.75)°、(1.31±0.65)°、(1.00±0.60)°;试验组(0.07±0.08)cm、(0.14±0.13)cm、(0.09±0.09)cm、(0.73±0.62)°、(0.72±0.58)°、(0.66±0.52)°。两组数据使用曼恩-惠特尼U检验,Lat、Vrt、Pitch、Roll、Rtn方向有统计学差异(P<0.05),Lng方向无统计学差异(P>0.05)。结论:相对于单面网,双面网固定方式在头部立体定向放射治疗中减小了患者的摆位误差,提高了治疗精度,可使患者获益。Objective To compare the differences of setup accuracy in stereotactic body radiotherapy for head cancer using two different immobilization techniques,namely single shell positioning system(SSPS)and double shell positioning system(DSPS).Methods Forty patients treated with stereotactic body radiotherapy for head cancer were diveided into two groups and then immobilized with SSPS(control group)and DSPS(experimental group).After that the pre-treatment cone beam CT images were obtained,the setup errors were acquired through image registration based on bony landmarks for statistical analysis.Results The setup errors(Mean±SD)in lateral(Lat),longitudinal(Lng),vertical(Vrt),Pitch,Roll and rotation(Rtn)directions in control group were(0.26±0.15)cm,(0.13±0.10)cm,(0.19±0.13)cm,(1.40±0.75)°,(1.31±0.65)°and(1.00±0.60)°,respectively,and those in experimental group were(0.07±0.08)cm,(0.14±0.13)cm,(0.09±0.09)cm,(0.73±0.62)°,(0.72±0.58)°and(0.66±0.52)°,respectively.The results of Mann-Whitney U test showed that there were statistical differences between two gourps in Lat,Vrt,Pitch,Roll and Rtn directions(P<0.05),but not in Lng direction(P>0.05).Conclusion Compared with SSPS,DSPS reduces the setup errors in stereotactic body radiotherapy for head caner and improves treatment accuracy,thereby bringing benefits to patients.
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:3.144.127.26