作 者:艾海明[1] 胡敏露[2,3] 刘洪运 于晓华[2,3] 王国静 王卫东 AI Haiming;HU Minlu;LIU Hongyun;YU Xiaohua;WANG Guojing;WANG Weidong(Faculty of Science and Technology,Beijing Open University,Beijing 100081,China;Center of Medical Device R&D and Clinical Evaluation,Chinese PLA General Hospital,Beijing 100853,China;Key Laboratory of Biomedical Engineering and Translational Medicine,Ministry of Industry&Information Technology,Beijing 100853,China)
机构地区:[1]北京开放大学科学技术学院,北京100081 [2]中国人民解放军总医院医疗器械研发与临床评价中心,北京100853 [3]生物医学工程与转化医学国家工信部重点实验室,北京100853
出 处:《传感器与微系统》2020年第5期94-96,99,共4页Transducer and Microsystem Technologies
基 金:国家科技部重点研发计划资助项目(2013BAI03B03,2015BAI01B08);国家自然科学基金资助项目(61372047,61701540);北京市教委科技项目(KM201951160001)。
摘 要:基于黑体热辐射原理采用MLX90615-DAA专用医用温度测量探头和MSP430主控制器进行温度源数据采集和处理,并应用矫正查表法和两次线性回归校准算法分别计算出温度值和校准温度值,样机和黑体标准器同步进行71例测温对比实验。样机实测温度值(Lab_tTr)和黑体标准器所取温度值(Std_tTr)的均方根误差为0.085℃,相关系数为0.9995(p<0.001);Bland-Altman图显示Lab_tTr与Std_tTr平均偏倚绝对值为0.0069℃,98.6%(70/71)的点在95%一致性界限内。实验表明:本样机与黑体标准器测温结果具有较好的一致性。The design uses special-purpose medical temperature-measurement probe MLX90615-DAA and main controller MSP430 for temperature source data collection and processing based on the principle of blackbody thermal radiation,and correction lookup algorithm and twice linear regression calibration algorithm is respectively used for temperature value calculation and calibration,71 contrast experiments of simultaneous temperature-measurement are conducted by using the prototype and blackbody standard.The root mean square error between temperature-measurement result of the prototype(Lab_tTr)and value-setting result of blackbody standard(Std_tTr)is 0.085℃,and correlation coefficient is 0.9995(p<0.001).Bland-Altman plot show that the mean absolute bias of Lab_tTr and Std_tTr is 0.0069℃,98.6%(70/71)points are within 95%confidence interval of agreement limits.The experimental results show that the temperature-measurement results by the prototype and the blackbody standard are in good agreement.
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.201