流动注射分析法同时测定水中挥发酚及氰化物  被引量:2

Simultaneous Determination of Volatile Phenol and Cyanide in Water by Flow Injection Analysis

在线阅读下载全文

作  者:黄贤伟 Huang Xianwei(Huizhou Huiyang Environmental Monitoring Station,Huizhou 516000,China)

机构地区:[1]惠州市惠阳区环境监测站,广东惠州516000

出  处:《广东化工》2020年第9期183-185,共3页Guangdong Chemical Industry

摘  要:本文利用SKALAR SAN++连续流动分析仪对水中挥发酚及氰化物进行同时测定。实验结果表明,该方法对于挥发酚和氰化物,分别在0~0.200 mg/L和0~0.100 mg/L范围具有良好的线性关系,相关系数分别达到0.9999及0.9998;方法检出限分别为0.0007 mg/L及0.001 mg/L。实际水样加标回收率分别为95.0%~102%及96.4%~106%。该方法具有测试准确、分析快速,操作便捷,试剂消耗小等优点,适合于水质监测批量样品的测定分析。In this work,SKALAR SAN++continuous flow analysis was developed for the volatile phenol and cyanide in water.The results showed that the good linear relationships were obtained when the concentration range of 0~0.200 mg/L for volatile phenol and concentration range of 0~0.100 mg/L for the cyanide,and with correlation coefficient of 0.9999 and 0.9998,respectively.The detection limits were 0.0007 mg/L(volatile phenol)and 0.001 mg/L(cyanide),respectively.The recovery efficiency of water samples is 95.0%~102%and 96.4%~106%,respectively.The method has the advantages of accuracy,rapid analysis,convenient operation and low reagent consumption,which is suitable for analysis of batch water samples for the water quality monitoring.

关 键 词:SKALAR SAN++ 流动注射 挥发酚 氰化物 

分 类 号:O65[理学—分析化学]

 