检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
作 者:陈云 黄荷茜 CHEN Yun;HUANG Hexi(Department of Equipment,The Third Affiliated Hospital of Wenzhou Medical College,Wenzhou Zhejiang 325200,China)
机构地区:[1]温州医科大学附属第三医院设备科,浙江温州325200
出 处:《中国医疗设备》2020年第7期169-171,175,共4页China Medical Devices
摘 要:费森尤斯4008系列血透机在临床上被广泛应用,易发生各类流量方面故障,但目前还缺少对整个流量系统故障的分析和解决方案。本文首先介绍费森尤斯血透机流量系统的基本工作原理,然后收集常见流量方面的故障现象,结合水路图及实物图找出原因并解决问题。本故障检修可以为维修工程师提供参考。Fresenius 4008 series hemodialysis is widely used in clinical practice,all kinds of flow problems are easy to occur.But it still lack of analysis and solution for the faults of whole flow system.First the basic working principles of the flow system of fresenious hemodialys was introduced,then common faults of the flow system were collected.By combining the waterway map and physical map,we found out the causes and solutions.This troubleshooting can provide reference for maintenance engineers.
分 类 号:R197.39[医药卫生—卫生事业管理] TH789[医药卫生—公共卫生与预防医学]