“移动战伤单元”16排移动CT头部扫描试验报告  被引量:2

Head scanning report with 16 row mobile CT in“mobile War Wound Unit”

在线阅读下载全文

作  者:王松叶 易蕊[2] 段利达 张震 侯国峰 张志强[2] 王振方 刘凌华 于强 栗志利 荆亚新 代秋声 徐如祥[2] Wang Songye;Yi Rui;Duan Lida;Zhang Zhen;Hou Guofeng;Zhang Zhiqiang;Wang Zhenfang;Liu Linghua;Yu Qiang;Li Zhili;Jing Yaxin;Dai Qiusheng;Xu Ruxiang(The 66242 Army Hospital of PLA,Sonid Right Banner 011299,China;Department of Neurosurgery,The Seventh Medical Center,Chinese PLA General Hospital,Beijing 100700,China;The Health Company of 66019 PLA,Sonid Right Banner 011299,China;Jiangsu Mokoto Medical Technology Co.Ltd.,Suzhou 215123,China;BECHOICE(Beijing)Science and Technology Deve Lopment Ltd.,Beijing 100050,China)

机构地区:[1]66242部队医院,内蒙古苏尼特右旗011299 [2]解放军总医院第七医学中心神经外科,北京100700 [3]66019部队卫生连,内蒙古苏尼特右旗011299 [4]江苏摩科特医疗科技有限公司,苏州215123 [5]北创动力(北京)科技发展有限公司,北京100050

出  处:《中华神经创伤外科电子杂志》2020年第4期214-218,共5页Chinese Journal Of Neurotraumatic Surgery:Electronic Edition

基  金:全军十二五军事医学重点课题(BWS12J010);国家自然基金青年科学基金(81100971)。

摘  要:目的探讨国产16排移动CT在救护车运载条件下进行头部扫描成像的可行性。方法选取陆军某部训练基地的志愿者健康官兵95人,随机分为3组:Ⅰ组(77人),救护车在停止行驶状态下进行头部扫描;Ⅱ组(11人),救护车在行驶状态下进行头部扫描;Ⅲ组(7人),设置电磁波干扰环境。分析对比3组受检者在不同状态下,移动CT扫描成像质量、运动伪影、数据采集及信息传输情况。结果骨窗成像:3组受检者颅骨、眼眶、鼻蝶窦等成像清晰。脑组织窗成像:3组受检者均可清晰显示眼球、视神经、脑干、脑皮质及脑白质等结构,但颅底部分层面均存在不同程度的运动伪影,其中Ⅰ组16.88%(13/77)受检者有较轻微的线状运动伪影,Ⅱ组81.82%(9/11)受检者有较明显的运动伪影;Ⅲ组14.29%(1/7)受检者有轻微线状运动伪影。经χ^2检验,Ⅰ组、Ⅲ组分别与Ⅱ组比较,差异均有统计学意义(χ^2=21.645、7.901,均P<0.05);Ⅰ组与Ⅲ组比较,差异无统计学意义(χ^2=0.031,P>0.05)。在电磁干扰状态下,移动CT数据采集及信息传输正常,3组辐射剂量CTDIvol均为36.27 mGy。结论16排移动CT在救护车行驶及电磁干扰状态下扫描成像稳定安全,数据采集及信息传输正常。Objective To explore the feasibility of head scanning with 16 row mobile CT made in China under the condition of ambulance transportation.Methods Ninety-five healthy officers and soldiers(volunteers)from a training base of the army were randomly divided into 3 groups:groupⅠ(n=77),ambulance head scanning in the stop state;groupⅡ(n=11),ambulance head scanning in driving state;groupⅢ(n=7),setting electromagnetic interference environment.The imaging quality,motion artifact,data acquisition and information transmission of the 3 groups in different states were analyzed and compared.Results Bone window imaging:The skull,orbit and sphenoid sinus in group were clearly imaged in three groups.Window imaging of brain tissue:The structures of eyeball,optic nerve,brain stem,cortex and white matter were clearly displayed in 3 groups,but there were different degrees of motion artifacts in the skull base.Among them,16.88%(13/77)subjects had slight linear motion artifacts in groupⅠ,81.82%(9/11)subjects had obvious motion artifacts in groupⅡ,14.29%(1/7)subjects had slight linear motion artifacts in groupⅢ.χ^2 test showed that there were significant differences in between groupⅠ,groupⅢand groupⅡ(χ^2=21.645,7.901,both P<0.05);there was no significant difference between group III and groupⅠ(χ^2=0.031,P>0.05).The data acquisition and information transmission of mobile CT are normal in the state of electromagnetic interference.The dose of CTDIvol was 36.27 mGy in 3 groups.Conclusion The 16 row mobile CT scan image is stable and safe under the conditions of ambulance driving and electromagnetic interference,and the data are normal of acquisition and information transmission.

关 键 词:移动CT CT扫描 头部成像 救护车 

分 类 号:R82[医药卫生—临床医学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象