作 者:康盛伟[1,2] 王培 祁国海[1,2] 刘操 龚岚[3] 黎杰 肖明勇[1,2] 唐斌[1,2] 郎锦义[1,2] KANG Shengwei;WANG Pei;QI Guohai;LIU Cao;GONG Lan;LI Jie;XIAO Mingyong;TANG Bin;LANG Jin-yi(Sichuan Cancer Hospital&Institute,Sichuan Cancer Center,Chengdu 610041,China;School of Medicine,University of Electronic Science and Technology of China,Chengdu 610041,China;National Institute of Measurement and testing technology,Chengdu,610000,China)
机构地区:[1]四川省肿瘤医院·研究所/四川省癌症防治中心,四川成都610041 [2]电子科技大学医学院,四川成都610041 [3]中国测试技术研究院,四川成都610000
出 处:《中国医学物理学杂志》2020年第8期945-950,共6页Chinese Journal of Medical Physics
基 金:四川省卫计委科研课题(17PJ585);国家重大科学仪器设备开发专项(2013YQ090811)。
摘 要:目的:对新型国产二维矩阵剂量验证系统在临床条件下进行测试,检验其是否能够满足临床使用需要。方法:参照GB15213-94对用来检测国产二维矩阵剂量验证系统的医用直线加速器进行检测调整,使其达到国家标准。使用新型国产二维矩阵剂量验证系统,对标准照射野下的绝对剂量重复性,标准照射野下的剂量线性,平坦度、对称性,真实病例放疗计划验证进行测试。结果:标准照射野下的绝对剂量重复性检测,其变异系数小于0.7%,符合测试要求;标准照射野下的剂量线性检测与电离室检测结果相比,无明显差异;平坦度检测±3%以内、对称性检测±2%以内,均满足临床使用要求;真实病例计划验证γ通过率均大于98%,完全满足临床放疗计划验证要求。结论:新型国产二维矩阵剂量验证系统具备点剂量、面剂量测量功能,能够对加速器基本剂量性能进行检测,达到临床使用要求;能够实现放疗计划系统的DICOM数据导入,与实际测量结果比较分析,达到临床计划验证要求。Objective To test a new type of domestic two-dimensional(2D)matrix dosimetric verification system in the clinical condition for evaluating whether the system can meet the clinical requirements.Methods The medical linear accelerator which was used to test the domestic 2D matrix dosimetric verification system was firstly tested and adjusted according to GB15213-94,thus ensuring that the accelerator met the national standard.Then the new type of domestic 2D matrix dose verification system is used for testing the repeatability of absolute dose measurement and dose linearity in the standard radiation field,flatness and symmetry,and validating actual radiotherapy plans.Results The test on the repeatability of absolute dose measurement in the standard radiation field showed that the coefficient of variation was less than 0.7%,which was consistent with the test requirements.There was no significant difference in the test results of dose linearity in the standard radiation field between 2D matrix dosimetric verification system and ionization chamber.The test also revealed that both flatness(within±3%)and symmetry(within±2%)met the clinical requirements.Moreover,the validation of actual treatment plans showed that the gamma passing rates were all above 98%,which fully satisfied the clinical requirements.Conclusion The novel domestic 2D matrix dosimetric system can be used to not only measure point and planar doses for the evaluation of the basic dose performances of the accelerator,but also realize the import of the DICOM data exported from the treatment planning system which are then compared with actual measurements,thereby meeting the clinical plan verification requirements.
分 类 号:R318[医药卫生—生物医学工程]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.127