彩色多普勒超声在血液透析患者自体动静脉内瘘血流量测量中的应用  被引量:22

Color Doppler Ultrasonography for the Measurement of Access Blood Flow in Hemodialysis Patients with Native Arteriovenous Fistula

在线阅读下载全文

作  者:段秀芳 郑茜子[1] 刘瑶[1] 王秀云[2] 邱素真 金其庄[1] Duan Xiufang;Zheng Xizi;Liu Yao;Wang Xiuyun;Qiu Suzhen;Jin Qizhuang(Department of Nephrology,the First Hospital,Institute of Nephrology,Peking University,Key Lab of Renal Disease,Ministry of Health of China,Beijing 100034,China;Department of Abdominal Ultrasound,the Fist Affiliqted Hospital of Harbin Medical University,Harbin,Heilongjiang 150001,China;Department of Nephrology,Henan Shangcai People′s Hospital,Shangcai,Henan 463800,China)

机构地区:[1]北京大学第一医院肾内科,北京大学肾脏病研究所,卫生部肾脏疾病重点实验室,北京市100034 [2]哈尔滨医科大学附属第一医院腹部超声科 [3]河南省上蔡县人民医院肾内科

出  处:《中国超声医学杂志》2020年第9期833-835,共3页Chinese Journal of Ultrasound in Medicine

基  金:北京大学第一医院科研种子基金(No.2019SF28)。

摘  要:目的探讨彩色多普勒超声在自体动静脉内瘘血流量测量中的应用价值。方法彩色多普勒超声测量自体动静脉内瘘血流量,以超声稀释法作为金标准,采用Pearson线性相关比较其一致性。结果共103例低位桡动脉-头静脉吻合的自体动静脉内瘘患者纳入本研究,超声稀释法测量内瘘流量(Qa)为(955.1±441.0)ml/min,彩色多普勒超声测量头静脉流量为(914.2±468.5)ml/min,肱动脉流量为(915.8±405.9)ml/min。所有患者中,肱动脉流量与头静脉流量均与Qa显著相关(P<0.001),且肱动脉相关系数R高于头静脉(0.856 vs 0.808)。在鼻烟窝、腕部、前臂等不同部位亚组中,肱动脉相关系数R均高于头静脉。在11例(10.7%)高流量内瘘中,头静脉流量与Qa相关性(R=0.852)高于肱动脉(R=0.773)。结论彩色多普勒超声可用于自体动静脉内瘘血流量监测,测量部位首选肱动脉,但是在高流量内瘘中,测量部位可选择头静脉。Objective To evaluate the clinical value of color Doppler ultrasonography for the measurement of access blood flow in hemodialysis patients with native arteriovenous fistula.Methods Color Doppler ultrasound was performed to measure access blood flow,using the ultrasound dilution method as the gold standard.Pearson linear correlation was used to evaluate the consistency between the two methods.Results 103 patients were enrolled in the study.The mean values of blood flow obtained with color doppler ultrasound were(914.2±468.5)ml/min in the cephalic vein,(915.8±405.9)ml/min in the brachial artery.Both were significantly correlated with blood flow obtained with ultrasound dilution[Qa(955.1±441.0)ml/min,both P<0.001]Blood flow in the brachial artery showed a greater correlation with Qa than the cephalic vein in all patients(R=0.856 vs 0.808),and in different anastomosis subgroups.For 11 patients with high flow fistulas(10.7%),blood flow measured in the cephalic vein showed a greater correlation(R=0.852)than the brachial artery(R=0.773).Conclusions Color Doppler ultrasound represents a good indirect method for blood flow measurement of native arteriovenous fistula.In general,brachial artery is more suitable for ultrasound measurement than cephalic vein except for high blood flow fistulas.

关 键 词:彩色多普勒超声 肾透析 动静脉内瘘 通路血流量 

分 类 号:R445.1[医药卫生—影像医学与核医学] R692.5[医药卫生—诊断学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象