检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
作 者:周崇兴[1] 梁君 张辉[3] 肖瑶[3] ZHOU Chong-xing;LIANG Jun;ZHANG Hui(Department of AIDS Prevention and Control,Guangxi Zhaung Autonomous Region Center for Disease Control and Prevention,Nanning,Guangxi Zhaung Autonomous Region 530028,China;不详)
机构地区:[1]广西壮族自治区疾病预防控制中心艾滋病防制所,南宁530028 [2]北京市通州区中心血站 [3]中国疾病预防控制中心性病艾滋病预防控制中心
出 处:《中国公共卫生》2020年第9期1361-1364,共4页Chinese Journal of Public Health
基 金:广西壮族自治区疾病预防控制中心青年基金(GXCDC201506)。
摘 要:目的获得国产通用型CD4^+T淋巴细胞(CD4)检测试剂应用于开放型流式细胞仪检测结果的准确性。方法于2014年7月1-30日对艾滋病病毒(HIV)感染者样本全血样本206份进行平行检测,以BD 4色试剂配套BD FACSCalibur流式细胞仪检测结果为参考,评价同生3色(CD3/8/4、CD3/4/45)、4色试剂(CD3/8/45/4)在ACEA NovoCyte流式细胞仪获得T淋巴细胞亚群绝对数的准确性。结果206份全血样本分析结果如下:同生试剂各参数与参考方法相关系数均>0.96。同生3色(CD3/8/4)、(CD3/4/45)和4色(CD3/8/45/4)检测CD4绝对数的平均偏差分别为(2.49±50.5)、(–3.1±43.8)和(–6.2±58.2)个/μL,与参考方法相比均无统计学差异;CD3绝对数的平均偏差分别为(10.3±115.3)、(–1.7±108.0)和(14.3±125.4)个/μL,均无统计学差异;CD8绝对数平均偏差分别为(–4.6±78.6)和(–19.6±87.2)个/μL。结论同生试剂应用于ACEA NovoCyte流式细胞仪均具有良好的准确性,可用于常规T淋巴细胞检测。Objective To evaluate the accuracy of a domestic reagent(Tongsheng reagent)for determination of absolute count of cluster of differentiation 4 positive(CD4^+)T lymphocytes with different multicolor flow cytometry.Methods Parallel tests for accuracy in absolute counts of T lymphocyte subgroups detected using domestic reagents and instrument were performed on whole blood samples of 206 human immunodeficiency virus(HIV)infected people during July 2014.The domestic reagents used in the detection were Tongsheng three color(CD3-FITC/CD8-PE/CD4-APC[CD3/8/4]/CD3-FITC/CD4-PE/CD45-PerCP[CD3/4/45])and Tongsheng four color(CD3-FITC/CD8-PE/CD45-PerCP/CD4-APC[CD3/8/45/4])and the domestic instrument was ACEA NovoCyte flow cytometry.The reference reagent and instrument were BD four color reagent kit and ral BD FACSCalibur flow cytometry.Results For all the blood samples,the correlation coefficients between the results from detections with domestic reagents and instrument and reference reagent and instrument were all greater than 0.96.Compared with the results using reference reagent and instrument,when using Tongsheng three color(CD3/8/4 or CD3/4/45)and Tongsheng four color(CD3/8/45/4)reagents and ACEA NovoCyte flow cytometry,the mean deviations of absolute counts(cell/μL)were 2.49±50.5,–3.1±43.8 and–6.2±58.2 for CD4;10.3±115.3,–1.7±108.0 and 14.3±125.4 for CD3;–4.6±78.6 and–19.6±87.2 for CD8,respectively.No significant differences were observed between the absolute counts of CD4,CD3,and CD8 T lymphocytes determined with domestic and reference reagents and instrument.Conclusion Domestic Tongsheng three color and four color reagents and ACEA NovoCyte can be used for accurate determinations of T lymphocytes absolute counts.
分 类 号:R331[医药卫生—人体生理学]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:3.149.253.148