作 者:李伟 谭超 LI Wei;TAN Chao(Department of Equipment,Yancheng No.1 People's Hospital,Yancheng 224001,China;不详)
机构地区:[1]盐城市第一人民医院设备处,江苏盐城224001
出 处:《中国医学装备》2020年第11期200-202,共3页China Medical Equipment
摘 要:分析Synergy医用电子直线加速器Rest Motor联锁电路原理及相关故障,讨论Rest Motor联锁工作所需要的基本条件和导致其联锁故障的可能原因及诊断方法,当出现RestMotor联锁且无法消除时,结合其相关电路分析可能的故障原因,利用观察、测量以及排除等方法找出故障原因,进行维修。To analyze the principle and related faults of interlock circuit of Rest Motor of Synergy medical electronic linear accelerator,and introduce the required basic conditions of Rest Motor interlock work,and the possible reasons that caused the interlock fault and diagnostic methods of interlock faults.When the equipment occurred Rest Motor interlock and couldn’t eliminate this fault,the possible fault reasons were analyzed by combining with relevant circuit.And the fault reasons were found by using observation,measurement,elimination and other means so as to provide reference for the maintenance of linear accelerator.