检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
作 者:刘毅 杜延春 许鹏 李世君 刘媛 夏华 LIU Yi;DU Yanchun;XU Peng;LI Shijun;LIU Yuan;XIA Hua(Shandong Institute of Metrology,Jinan 250014,China)
出 处:《计量科学与技术》2020年第10期44-46,59,共4页Metrology Science and Technology
摘 要:分析了常见的50Ω/50μH+5Ω和50Ω/5μH+1Ω两种不同V型人工电源网络的差异,对直流电能表分别使用这两种电源网络进行端子传导骚扰的测量,在典型频率点的测量结果进行了分析比较。对直流电能表的端子传导骚扰测量来讲,50Ω/50μH+5Ω网络可用作摸底性测试,而标准符合性测量结果必须以50Ω/5μH+1Ω网络为准。The differences between two different V-type artificial mains networks,50Ω/50μH + 5Ω and 50Ω/5μH +1Ω,are analyzed and the measurements of terminal conducted turbulence at typical frequency points are compared for the DC power meter using each of the two artificial mains networks. The 50Ω/50μH + 5Ω network can be used as the baseline test for the terminal conduction disturbance measurement of DC power meter,while the standard compliance measurement result must be based on the 50Ω/5μH + 1Ω network.
分 类 号:TM933.4[电气工程—电力电子与电力传动]
正在载入数据...