医学工程保障中除颤监护仪质量检测分析研究  被引量:4

Study on the Quality Detection of Defibrillation Monitor During Medical Engineering Support

在线阅读下载全文

作  者:晋学松 万振宽[1] JIN Xuesong;WAN Zhenkuan(Department of Medical Engineering,Beijing Anding Hospital of Capital Medical University,Beijing 100088,China)

机构地区:[1]首都医科大学附属北京安定医院医学工程处,北京100088

出  处:《中国医疗设备》2020年第12期67-69,74,共4页China Medical Devices

摘  要:目的通过质量控制,检测两种除颤监护仪的精神科医院临床应用效能。方法采用FLUKE除颤分析仪对本院同时采购的两个品牌共16台除颤监护器进行包括外观和正常性检测、释放能量准确性、充电时间、同步延时时间、心率示值等5个项目的质控检测。结果除1台美国A除颤器释放能量检测不合格,其他合格率均为100%。结论上海B除颤仪在释放能量准确性、充电时间等两个方面都要比美国A除颤监护仪稳定、有效、安全,更适合精神专科医院临床使用。Objective To test the clinical application efficacy of two defibrillation monitors in psychiatric hospital through quality control.Methods Using the FLUKE defibrillation analyzer to test 16 defibrillation monitors from two brands,which are purchased by our hospital at the same time,including monitors’appearances and normal detection,release of energy accuracy,charging time,synchronization delay time and heart rate indication.Results Except one American A defibrillator failed the energy release test,the others are all 100%.Conclusion By comparing and analyzing the results of quality control test and combining the experience of psychiatric clinic,Shanghai B defibrillator is more stable,effective and safer than American A defibrillation monitor in terms of accuracy of energy release and charging time,meaning it is more suitable for psychiatric hospital clinic use.

关 键 词:除颤监护仪 质量控制 急救设备 

分 类 号:R197.39[医药卫生—卫生事业管理]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象