铜和铜-铝复合滤过板对数字X线摄影透射线率影响的模体实验  被引量:2

Model Experiment of Effect of Copper and Copper-Aluminum Composite Filter Plate on Transmission Rate of Digital X-Ray Photography

在线阅读下载全文

作  者:武卫平 殷硕 刘建新 WU Weiping;YIN Shuo;LIU Jianxin(Department of Radiology,Emergency General Hospital,Beijing 100028,China)

机构地区:[1]应急总医院影像科,北京100028

出  处:《中国医疗设备》2020年第12期99-102,108,共5页China Medical Devices

摘  要:目的比较铜和铜-铝复合滤过板对X线的透射线率,为合理选用附加滤过板提供依据。方法Definium6000数字化X线摄影机,70~130 kV,200 mA,AEC模式,源-像距150 cm和180 cm。三组附加滤过分别为0.1 mm Cu和0.1 mm Cu+1.5 mm Al、0.2 mm Cu和0.2 mm Cu+1.5 mm Al、0.3 mm Cu和0.3 mm Cu+1.5 mm Al。对18 cm厚聚乙烯模体曝光,检测入射剂量、透射剂量,计算透射剂量率。使用SPSS17.0软件对铜和相同厚度铜-铝复合滤过板的入射剂量、透射剂量、以及透射剂量率进行比较。结果源-像距为150 cm时,透射线率(%)中位数分别为7.34和7.87、8.60和9.07、9.64和10.13;源-像距为180 cm时,透射线率(%)中位数分别为7.33和7.97、8.66和9.26、9.84和10.27。铜滤过厚度相同时,铜-铝复合滤过的透射线率比单独使用铜滤过时大,差别有统计学意义(P<0.05)。以管电压为分组变量,不同厚度铜滤过与铜-铝复合滤过的透射剂量差别没有统计学意义(P>0.05)。结论使用厚度大的附加滤过可以增加X线透射线率,并且铜-铝复合滤过板的X线透射线率比单独使用铜滤过时相对大,有利于减少患者所受辐射剂量。Objective Comparing the X-ray transmission rate of copper and copper aluminum composite filter plates,so as to provide the basis for reasonable selection of additional filter plates.Methods Definium6000 digital X-ray camera,70-130 kV,200 mA,AEC mode,source-image distance 150 cm and 180 cm.The three additional groups of filters were 0.1 mm Cu and 0.1 mm Cu+1.5 mm Al,0.2 mm Cu and 0.2 mm Cu+1.5 mm Al,0.3 mm Cu and 0.3 mm Cu 1.5 mm Al,respectively.The 18 cm thick polyethylene phantom was exposed,the incident dose and transmission dose were detected,and the transmission dose rate was calculated.The incident dose,transmission dose,and transmission dose rate of copper and the same thickness copper-aluminum composite filter plate were compared using SPSS 17.0 software.Results When the source image distance was 150 cm,the median transmittance(%)was 7.34 and 7.87,8.60 and 9.07,9.64 and 10.13,respectively,Meanwhile,when the source image distance was 180 cm,the median transmittance(%)was 7.33 and 7.97,8.66 and 9.26,9.84 and 10.27,respectively.Based on the same thickness of copper filter,the ray transmission rate of copper aluminum composite filter was larger than that of single copper filter,and the difference was statistically significant(P<0.05).Moreover,taking tube voltage as group variable,there was no significant difference in transmission dose between different thickness copper filter and copper aluminum composite filter(P>0.05).Conclusion The X-ray transmission rate of copper aluminum composite filter plate is higher than that of single copper filter plate,which is beneficial to reduce the radiation dose of patients.

关 键 词:数字X射线摄影 附加滤过 辐射剂量 透射剂量率 

分 类 号:R144[医药卫生—公共卫生与预防医学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象