瓦里安Novalis Tx治疗床对放疗剂量的影响  被引量:3

Effect of a Treatment Couch of Varian Novalis Tx on Radiotherapy Dose

在线阅读下载全文

作  者:吴友涛 黎明 唐德文 李勇 张德康 WU Youtao;LI Ming;TANG Dewen;LI Yong;ZHANG Dekang(Department of Radiology,Sichuan Cancer Hospital Institute,Sichuan Cancer Center,School of Medicine,University of Electronic Science and Technology,Chengdu Sichuan 610041,China)

机构地区:[1]四川省肿瘤医院·研究所四川省癌症防治中心电子科技大学医学院放疗中心,四川成都610041

出  处:《中国医疗设备》2020年第S01期47-49,59,共4页China Medical Devices

基  金:四川省科技厅项目(2020YFS0393)。

摘  要:目的研究瓦里安Novalis Tx医用电子直线加速器碳纤维治疗床对放疗剂量的影响。方法采用等中心SCD100 cm照射,X射线能量为6 MV,每个射野机器跳数设置为100 MU,改变机臂角度,让高能X射线从不同角度穿透治疗床,使用电离室测量吸收剂量,计算出对应的衰减因子。结果从测量结果来看,床板连接处衰减因子最高达49.52%,主床板衰减因子最高达8.91%,延长板衰减因子最高达5.76%。差异有统计学意义(P<0.05)。结论在临床工作中照射野应避开床板连接处,TPS中应导入治疗床模型以便修正治疗床剂量衰减的影响。Objective To study the effect of carbon fiber couch of Novalis Tx medical electron linear accelerator on radiotherapy dose.Methods Isocentric SCD100 cm irradiation was used.The x-ray energy was 6 MV,the hop number of each field machine was set to 100 MU.The Angle of the machine arm was changed to allow high-energy X-rays to penetrate the treatment bed from different angles.The absorption dose was measured using the ionization chamber,and the corresponding attenuation factor was calculated.Results According to the measurement results,the attenuation factor of carbon fiber couch plate joint was up to 49.52%,the attenuation factor of the main couch plate was up to 8.91%,and the attenuation factor of the extended couch plate was up to 5.76%.The difference was statistically significant(P<0.05).Conclusion In clinical work,the exposure field should avoid the bed plate connection,and the treatment bed model should be introduced in TPS to correct the effect of dose attenuation on the treatment bed.

关 键 词:医用电子直线加速器 治疗床 衰减因子 放疗剂量 

分 类 号:R318.6[医药卫生—生物医学工程]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象