Clinac iX型医用直线加速器MOD联锁故障检修  被引量:2

Clinac iX linear accelerator MOD interlock fault maintenance

在线阅读下载全文

作  者:陈晓[1,2] 刘旭红[1] 赵永军[1] 王禅[1] CHEN Xiao;LIU Xu-hong;ZHAO Yong-jun(Radiation Therapy Center,The Tumor Hospital of Yunnan Province,Kunming 650118,China;不详)

机构地区:[1]云南省肿瘤医院放疗中心,云南昆明650118 [2]昆明医科大学第三附属医院放疗中心,云南昆明650118

出  处:《中国医学装备》2021年第2期178-179,共2页China Medical Equipment

摘  要:分析Clinac iX型医用直线加速器MOD联锁的发生原因,检查高压脉冲形成网络(PFN)放电电路,同时发现高压脉冲电缆漏电或速调管灯丝局部短路,均会引发MOD联锁,为此针对性地进行检修,使相关故障得到解决。Analyzed the causes of the MOD interlock of Clinac iX linear accelerator,inspectchecked the high-voltage pulse forming network(PFN)discharge circuit,and found that the leakage of the high voltage pulse cable leakage or the partial short circuit of klystron filament would cause the MOD interlock.The targeted maitenancemaintenance was carried out,and related faults were resolved,which providesd reference for maintenance personnel to repair similar equipment.

关 键 词:调制器 放电电路 脉冲电缆 油箱 速调管 联锁故障 

分 类 号:R812[医药卫生—放射医学]

 