检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
作 者:何流[1] 方宣城 申才良 李浩[1] 李瑜 HE Liu;FANG Xuan-cheng;SHEN Cai-liang;LI Hao;LI Yu(Department of Spine Surgery,the First People’s Hospital of Anqing,Anqing,Anhui 246003,China)
机构地区:[1]安庆市第一人民医院脊柱外科,安徽安庆246003
出 处:《中华全科医学》2021年第1期42-45,共4页Chinese Journal of General Practice
基 金:国家重点研发计划资助项目(2018YFC0114708)。
摘 要:目的探讨天玑骨科手术机器人系统辅助下置入胸腰椎椎弓根螺钉的临床效果。方法将2018年8月—2020年5月于安庆市第一人民医院脊柱外科接受胸腰椎椎弓根螺钉内固定手术的40例患者,根据手术方式分为机器人辅助手术组和"C"形臂透视下徒手组,2组各20例。记录每台手术椎弓根螺钉的置入时间及术中透视次数,根据术后X线和三维CT评价2组螺钉的位置情况(准确率),评判螺钉置入的准确性和安全性。结果机器人辅助手术组每枚螺钉置入时间为平均(13.85±1.41)min,"C"形臂X线透视下徒手置钉组每枚螺钉置入时间为平均(14.03±1.84)min,2组数据差异无统计学意义(P>0.05)。机器人辅助手术组共置入螺钉94枚,Ⅰ级螺钉90枚,Ⅱ级螺钉4枚,Ⅲ级螺钉0枚,准确率为95.74%(90/94);"C"形臂X线透视下徒手置钉组共置入螺钉96枚,Ⅰ级螺钉83枚,Ⅱ级螺钉10枚,Ⅲ级螺钉3枚,准确率为86.46%(83/96);2组数据对比,差异有统计学意义(P<0.05)。机器人辅助手术组术中的透视次数为(8.15±1.53)次,"C"形臂X线透视下徒手置钉组术中的透视次数为(13.35±2.58)次,2组比较差异有统计学意义(P<0.01)。结论天玑骨科手术机器人辅助手术能明显提高置入椎弓根螺钉的准确性及安全性,同时也减少了术中放射线辐射,但并未明显增加置钉时间,在脊柱手术中有着巨大的手术潜能和优势。Objective To explore the clinical effect of thoracic and lumbar pedicle screw placement with the assistance of a robotic system for orthopaedic surgery.Methods Total 40 patients underwent thoracolumbar pedicle screw internal fixation at the Department of Spine Surgery,Anqing First People’s Hospital From August 2018 to May 2020 were divided into two groups:robot assisted operation group and"C"arm fluoroscopy group with 20 cases in each group according to the operation methods.The placement time of each pedicle screw and the number of intraoperative fluoroscopy were recorded,and the position(accuracy)of the two groups of screws was evaluated on the basis of postoperative X-ray and three-dimensional CT.Moreover,the accuracy and safety of screw placement were evaluated.Results In the robot-assisted surgery group,the average screw insertion time was(13.85±1.41)min,and in the"C"arm X-ray fluoroscopy,the screw insertion time was(14.03±1.84)min,and the difference between the two groups of data was not statistically significant(P>0.05).A total of 94 screws were placed in the robot-assisted surgery group(90 in ClassⅠscrews,4 in ClassⅡscrews and 0 in ClassⅢscrews),with an accuracy rate of 95.74%(90/94).Using the"C"arm X-ray fluoroscopy,a total of 96 screws(83 GradeⅠscrews,10 GradeⅡscrews and 3 GradeⅢscrews)were placed in the nail placement group,with an accuracy rate of 86.46%(83/96).The difference between the two groups was significant(P<0.05).The number of fluoroscopy in the robot-assisted surgery group was(8.15±1.53)times,and the number of fluoroscopy in the"C"arm X-ray fluoroscopy was(13.35±2.58)times.The difference between the two groups was statistically significant(P<0.05).Conclusion The robot-assisted Tianji orthopaedic surgery can improve the accuracy and safety of pedicle screw placement and reduce intraoperative radiation,but it does not significantly increase the time of nail placement.Furthermore,it has a great application potential in spinal surgery.
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.15