作  者:陈诚 张疆华[1] 邢强[1] 杨徐 周贤惠[1] 李耀东[1] 芦颜美[1] 祖克拉·吐尔洪 张玲[1] 汤宝鹏[1] Chen Cheng;Zhang Jianghua;Xing Qiang;Yang Xu;Zhou Xianhui;Li Yaodong;Lu Yanmei;Zukela·Tuerhong;Zhang Ling;Tang Baopeng(Department of Pacing and Electrophysiology,Department of Cardiac Electrophysiology and Remodeling,The First Affiliated Hospital of Xinjiang Medical University,Urumqi 830054,China)

机构地区:[1]新疆医科大学第一附属医院心脏起搏电生理科,新疆心电生理与心脏重塑重点实验室,乌鲁木齐830054

出  处:《中华心律失常学杂志》2021年第1期37-42,共6页Chinese Journal of Cardiac Arrhythmias

摘  要:目的探究右心室造影对无导线起搏器植入位置的指引作用,评价植入后的安全性和有效性,比较与传统单腔起搏器植入术式的优劣。方法本研究为前瞻性、观察性研究,选择2019年12月至2020年12月期间新疆医科大学第一附属医院60例行无导线起搏器植入的患者,根据无导线起搏器是否在释放前行右心室造影分为造影组及非造影组,并均于术后行心脏三维重建CT,比较无导线起搏器释放的部位差异。随访期间收集起搏器植入术中、术后的阈值、感知、阻抗等参数和不良事件发生情况;与传统单腔起搏器植入术比较手术时间、曝光量等。结果60例符合适应证的患者均成功植入无导线起搏器(Micra,美国美敦力公司)。造影组40例,非造影组20例,最后均通过心脏三维重建CT确定无导线起搏器的植入部位。非造影组中11例(55.0%,11/20)的患者植入低位间隔,造影组中33例(33/40,82.5%)患者植入中位间隔,差异具有统计学意义(P<0.05)。平均随访(105±30)d,阈值、感知随时间推移逐渐趋于稳定,阻抗进一步下降1个月后逐渐趋于稳定。除1例患者术区发生轻微血肿外,其余患者无植入器械相关并发症发生。前10例无导线起搏器植入手术时间[(60.00±9.28)min]、X线曝光时间[(14.90±2.28)min]、X线曝光量[(902.50±150.94)mGy]均明显高于同时期传统起搏器植入术,差异有统计学意义(P<0.05)。随着术者熟练程度的增加,上述指标逐渐缩短并趋于稳定。后20例无导线起搏器与传统起搏器手术时间、X线曝光时间差异无统计学意义,但X线曝光量明显大于传统起搏器,差异具有统计学意义[(564.00±115.90)mGy对(65.25±19.94)mGy,P<0.05)。结论本研究发现通过右心室造影可以增加无导线起搏器的精准定位。早期随访临床应用安全、有效。与传统单腔起搏器植入术式比较,虽曝光量增加,但手术更加微创,安全性更优。Objective To assess the guiding role of right ventriculography in the placement of leadless pacemaker.To evaluate the safety and efficacy of leadless pacemaker and to compare the advantages and disadvantages with traditional single-chamber pacemaker.Methods In this prospective and observational study,a total of 60 patients with Micra leadless pacemaker implantation from December 2019 to December 2020 in The First Affiliated Hospital of Xinjiang Medical University were divided into right ventriculography group(40 cases)and non-right ventriculography group(20 cases).All the cases were verified by the cardiac three-dimensional reconstruction CT scan.The pacemaker threshold,sensing amplitude,impedance and adverse events were collected,and the operation time and radiation exposure were compared with the traditional single chamber pacemaker.Results Micra leadless pacemakers were successfully implanted in all the 60 patients with indications.In the ventriculography group,there were 33 cases(82.5%)in the middle septum.In the non-ventriculography group,there were 11 cases(55.0%)in the low septum.There was statistical significance(P<0.05)in the septum rate between the two groups.During the follow-up period,threshold and impedance were decreased,sensing amplitude was increased.Except for one patient had hematoma in the operative area,there were no implantation-related complications in the other patients.There were no adverse events during the follow-up period.The average threshold and impedance were decreased,sensing amplitude was increased.The operation time[(60.00±9.28)min],radiation time[(14.90±2.28)min],and radiation exposure dose[(902.50±150.94)mGy]in the first 10 cases were higher than that in traditional pacemaker implantation.However,the operation and radiation time was decreased in the last 20 cases with exception of sustained higher radiation exposure dose[(564.00±115.90)mGy vs.(62.25±19.94)mGy].Conclusion Right ventriculography maybe an innovative and valid method to guide leadless pacemaker to the optimal p

关 键 词:心脏起搏器 人工 无导线起搏器 右心室造影 安全性 有效性 

分 类 号:R541.7[医药卫生—心血管疾病]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象