全自动生化分析仪检测血清脂蛋白a的分析性能评价  被引量:5

Evaluation of Analysis Performance of Automatic Biochemical Analyzer in Detecting Serum Lipoprotein A

在线阅读下载全文

作  者:薛云松[1] 李丽莉[1] 冯倩[1] 李宜铮[1] 周林华[1] 顾津伊 刘建梅 XUE Yun-song;LI Li-li;FENG Qian;LI Yi-zheng;ZHOU Lin-hua;GU Jin-yi;LIU Jian-mei(Dept.of Laboratory Medicine,The Second People's Hospital of Yunnan Province,Kunming Yunnan 650021,China)

机构地区:[1]云南省第二人民医院检验科,云南昆明650021

出  处:《昆明医科大学学报》2021年第3期49-53,共5页Journal of Kunming Medical University

基  金:云南省科技厅-昆明医科大学应用基础研究联合专项基金资助项目(2019FE001-092)。

摘  要:目的对在雅培C16000全自动生化分析仪上用胶乳凝集比浊法检测血清脂蛋白a[LP(a)]的分析性能进行评价,判断其是否能满足临床需求。方法参考国际、国内有关性能评价的文件和行业标准,结合工作实际从精密度、正确度、分析测量范围、临床可报告范围、抗干扰性能以及生物参考区间几个方面进行评价。结果LP(a)在质控水平1、2和混合血清样本中的批内不精密度分别为1.44%、0.45%、0.91%,批间不精密度分别为1.67%、0.81%、1.23%,都低于厂商声明的标准;与已通过ISO15189实验室认可的其他实验室相同检测系统进行样本的比对,相关系数r^(2)为0.9990;分析测量范围在11.0~919.0 mg/L之间;最大允许稀释倍数为10倍,最高可报告范围可扩展到9190.0 mg/L;采用商品化的干扰物质检测结果显示,游离胆红素20 mg/dL以下、结合胆红素20 mg/dL以下、乳糜浊度3000 FTU以下、血红蛋白500 mg/dL以下对LP(a)测定结果检测干扰的偏移量都小于10%,这些浓度的物质对LP(a)的检测结果干扰很小;以40例健康体检人群样本验证实验室引用的参考区间,验证通过。结论雅培全自动生化分析仪测定LP(a)的分析性能与厂家声明基本一致,能够满足临床需求,可以应用于临床。Objective To evaluate the assay performance of automatic biochemical analyzer in detecting serum lipoprotein A[LP(a)]by using latex agglutination-turbidimetry on a(Abbott C16000),so as to evaluate whether it can meet clinical requirements.Methods According to international and domestic performance evaluation documents and industry standards,the evaluation was carried out from the aspects of precision,accuracy a,analytical measurement range,clinical reportable range,anti-interference performance and biological reference range.Results The intra-batch imprecisions of LP(a)were 1.44%,0.45%and 0.91%at quality control level 1,2 and mixed serum samples,and the inter-batch imprecisions were 1.67%,0.81%and 1.23%,respectively,which were lower than the standard declared by the manufacturer.The correlation coefficient R^(2) was0.9990 when comparing the samples with the same testing system approved by ISO15189 laboratory.The analytical measurement range was 11.0-919.0 mg/L.The maximum allowable dilution was 10 times,and the maximum reportable range was extended to 9190.0 mg/L.Commercial interference substance detection results showed that the deviation of free bilirubin was less than 20 mg/dL,binding bilirubin was less than 20 mg/dL,chylous turbidity was less than 3000 FTU,hemoglobin was less than 500 mg/dL on the detection result of LP(a)was less than 10%,indicating that these concentrations of substances had little interference on the detection result of LP(a).The reference interval cited by the laboratory was verified by 40 healthy population samples.Conclusion The analytical performance of automatic biochemical analyzer in detecting LP(A)is basically consistent with the manufacturer’s statement,which can meet the clinical requirements and can be applied in clinical practice.

关 键 词:脂蛋白A 全自动生化分析仪 性能评价 

分 类 号:R331[医药卫生—人体生理学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象