作 者:刘丽 陆爽 敖丽萍 刘军廷[2] 程红[2] 黄玮浩 杨轮 张广川 米杰 杨翌 Liu Li;Lu Shuang;Ao Liping;Liu Junting;Cheng Hong;Huang Weihao;Yang Lun;Zhang Guangchuan;Mi Jie;Yang Yi(Department of Epidemiology and Biostatistics,School of Public Health,Guangdong Pharmaceutical University,Guangzhou 510006,China;Department of Epidemiology,Capital Institute of Pediatrics,Beijing 100020,China;Department of Non-communiccable Disease Management,Beijing Children's Hospital,Capital Medical University,National Center for Children's Health,Beijing 100045,China)
机构地区:[1]广东药科大学公共卫生学院流行病与卫生统计学系,广州510006 [2]首都儿科研究所流行病学研究室,北京100020 [3]国家儿童医学中心儿童慢病管理中心,首都医科大学附属北京儿童医院,北京100045
出 处:《中华流行病学杂志》2021年第3期475-481,共7页Chinese Journal of Epidemiology
基 金:国家自然科学基金(81874271,81973110)。
摘 要:目的评价生物电阻抗法(BIA)与双能X线吸收法(DXA)测量7~17岁儿童青少年体成分的一致性。方法对1431名儿童进行BIA和DXA检测。采用组内相关系数(ICC)和Bland-Altman分析评价方法间所测去脂体重和脂肪量的一致性。Bland-Altman分析在对数变换的数据中进行。结果男、女生中BIA与DXA所测去脂体重的ICC分别为0.986和0.974,脂肪量的ICC则分别为0.854和0.926。男生中BIA与DXA的去脂体重比值均值及一致性界限(LoA)分别为1.04和0.95~1.14,女生中则为1.02和0.90~1.15。男、女生中去脂体重的LoA范围均随年龄增长而变窄。男、女生中两种方法的脂肪量比值LoA范围均较宽,分别为0.40~1.27和0.48~1.48。此外,男、女生中均可看到各成分的比值LoA范围随BMI等级增高而变窄。结论BIA所测去脂体重与DXA的一致性良好,但BIA测量脂肪量的误差较大。肥胖儿童的BIA与DXA一致性优于消瘦和正常体重儿童。Objective To evaluate the consistency between bioelectrical impedance analysis(BIA)and dual-energy X-ray absorptiometry(DXA)in the measurement of body composition in children and adolescents aged 7-17 years.Methods Fat-free mass(FFM)and fat mass(FM)were measured by both BIA and DXA in 1431 children.The consistency between the methods was evaluated by intra-class correlation coefficients(ICCs)and Bland-Altman analysis.Logarithmic transformation of both measurements was performed before Bland-Altman analysis.Results The ICCs for FFM were 0.986 and 0.974 and ICCs for FM were 0.854 and 0.926 in boys and girls respectively.In boys,the mean ratio of FFMs by BIA and DXA was 1.04,with limits of Agreement(LoA)of 0.95-1.14,and in girls,the mean ratio of FFMs by BIA and DXA was 1.02,with the LoA of 0.90-1.15.The LoA of FFM became narrower with age in both boys and girls.Both boys and girls had the wide LoAs for FM(0.40-1.27 and 0.48-1.48,respectively).Additionally,the LoA ranges for FFM and FM narrowed with the increase of BMI level in both boys and girls.Conclusion For all children,BIA showed good consistency with DXA for FFM,whereas significant errors occurred in FM measurement.The consistency between BIA and DXA was better for obese children than for underweight or normal-weight children.
关 键 词:生物电阻抗法 双能X线吸收法 去脂体重 脂肪量 一致性评价
分 类 号:R179[医药卫生—妇幼卫生保健]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.249