高压氧舱内应用容量型输液泵的安全性和准确性研究  被引量:2

The safety and accuracy of using volumetric infusion pump in hyperbaric oxygen chamber

在线阅读下载全文

作  者:张蕾[1] 沙英[1] 何健[1] 王青[1] 刘亚玲[1] Zhang Lei;Sha Ying;He Jian;Wang Qing;Liu Yaling(Department of Hyperbaric Oxygen,Beijing Tiantan Hospital Affiliated to Capital Medical University,Beijing 100070,China)

机构地区:[1]首都医科大学附属北京天坛医院高压氧科,北京100070

出  处:《中华航海医学与高气压医学杂志》2021年第2期208-212,共5页Chinese Journal of Nautical Medicine and Hyperbaric Medicine

摘  要:目的观察高压氧舱内气压变化对容量型输液泵使用安全性和泵速准确性的影响。方法在高气压(表压0~0.10 MPa)和常压(北京地区大气压力)环境下,使用容量型输液泵泵速实时监测系统,测定输液泵设定流量为50、100 ml/h时的瞬时流量和平均流量,每种流量重复测定7次。观察容量型输液泵在高气压环境下(表压0~0.10 MPa)的使用情况,并比较在2种环境下输液泵的瞬时流量和平均流量。结果高压氧舱内匀速加、减压及高压稳压过程中,容量型输液泵可正常使用,未出现电器安全事故及使用故障。设定输液泵流量为50、100 ml/h时,输液泵的瞬时流量和平均流量在高气压(表压0~0.1 MPa)和常压环境下,测得结果差异无统计学意义(P>0.05),流量相对示值误差均在±2%之内。结论容量型输液泵可在高压氧舱内安全使用,且泵速准确。Objective To observe the influence of air pressure change in the hyperbaric oxygen chamber on the safety of the volumetric infusion pump operation and the accuracy of the flow rate.Methods The instantaneous flow rate and average flow rate of the volumetric infusion pump set at the flow rates of 50 and 100 ml/h were repeatedly measured for 7 times by the real-time flow rate monitoring system under high air pressure(gauge pressure 0-0.10 MPa)and normal pressure(atmospheric pressure in Beijing),respectively.The operation status of the volumetric infusion pump in the high-pressure environment(gauge pressure 0-0.10 MPa)was observed,and the instantaneous flow rates and average flow rates of the volumetric infusion pump under both high and normal pressures were compared.Results The volumetric infusion pump could operate smoothly,and no electrical safety accident or malfunction occurred in the scenarios of uniform pressurization and decompression and high-pressure stabilization in the hyperbaric oxygen chamber.When the flow rate of the infusion pump was set at 50 and 100 ml/h,comparing the instantaneous flow rate and average flow rate under the high pressure(gauge pressure 0-0.1 MPa)and the normal pressure environment respectively,there was no statistically significant difference(P>0.05),and the relative indication errors of the flow were all within±2%.Conclusion The volumetric infusion pump can safely operate and maintain accurate flow rate(50 ml/h and 100 ml/h)in hyperbaric oxygen chamber.

关 键 词:容量型输液泵 高压氧舱 安全性 准确性 

分 类 号:R197.39[医药卫生—卫生事业管理]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象